Introduction

Coronary artery disease (CAD), a leading cause of cardiovascular mortality in the world, is a disorder that occurs due to atherosclerotic occlusions of the coronary arteries. It is manifested by stable or unstable angina, myocardial infarction, sudden cardiac death and/or heart failure1,2. Single photon emission computed tomography (SPECT) myocardial perfusion imaging (MPI) is a well-established non-invasive diagnostic method to diagnose myocardial ischemia or infarctions in patients with known or suspected CAD3. SPECT MPI provides valuable information about the severity and extension of CAD, which helps in patient management4,5,6. With the advancement and the introduction of semi-conductor detector, cadmium zinc telluride (CZT) SPECT, the advantages are in the reduction of radiation doses, shorter acquisition time and improved overall detection efficiency7.

Technetium-99 m (Tc-99 m) is a gamma-emitting radioisotope with a principal photon energy of 140 keV and a physical half-life of 6.04 h. Tetrofosmin (Myoview™) is a pharmaceutical agent that is a lipophilic cationic diphosphine (trans-dioxo-bis) complex. When Tc-99 m pertechnetate (unlabelled Tc-99 m) is added to tetrofosmin in the presence of the reducing agent stannous ion, a lipophilic, cationic Tc-99 m tetrofosmin complex is formed8. Tc-99 m tetrofosmin is a cationic diphosphine with good myocardial uptake (first-pass extraction efficiency of ~ 50%) and retention with little redistribution from the myocardium over time9. Alternatively, radiotracers such as technetium-99 m radiolabelled with sestamibi or thallium-201 could be the other options for this study10.

The computed tomography (CT) coronary artery calcium score (CAC) is a non-invasive assessment used to evaluate the amount of coronary atherosclerosis. CAC can be used alone or alongside MPI, offering detailed anatomical insights into coronary health alongside MPI’s functional information. It helps further with risk stratification, assist in the assessment and guides management of patients with known or suspected CAD11. Integrating a CAC score with MPI provides complementary information in the form of coronary anatomical assessment alongside the presence or absence of ischemia12,13. The use of CAC is beneficial for assessing coronary artery disease risk, adding more critical information in MPI’s findings. However, it is crucial to consider the cumulative radiation dose to the patient when incorporating CAC into routine clinical practice.

Diagnostic reference level (DRL) and achievable dose (AD) are critical tools in medical imaging for ensuring patient radiation safety14,15. The DRL is defined as the 75th percentile of radiation exposure in regional or national surveys. DRL is used to identify situations where the doses may be unusually high. AD, representing the median or 50th percentile, is a newer standard aimed at representing doses realistically achievable by centres that are using standard techniques and commonly available equipment14,16. These benchmarks help imaging facilities evaluate and potentially adjust their protocols to optimize patient safety, with DRL acting as an upper threshold for review, and AD offering a broader target for routine practices. This study aims to establish a local baseline for DRL and AD in SPECT MPI at a single centre, thereby addressing the gaps in Malaysia’s current national DRL (NDRL), which was based on limited institutional data from 2005 to 2007. By analysing data within a single centre, this study aims to compare results with other studies to improve radiation safety and optimize local imaging practices. Moreover, establishing a locally relevant baseline can serve as a model for nationwide and international adoption of SPECT MPI procedures, contributing to the ongoing improvements in radiation protocols and providing valuable data for refining future NDRL and other international standards.

Methods

Study design

This study was approved by IJN Research Ethics Committee (project registration ID IJNREC/651/2024). It was a six-month retrospective study conducted from January 2024 to June 2024 by reviewing image data of adult patients referred for MPI study to this institution. Informed consent for each recorded data was waived by IJN Research Ethics Committee prior to study initiation. No major protocol changes occurred during the study period. Patients had MPI scan (either one-day or two-day protocol) using a cardiac-dedicated gamma cameras, followed by an external computed tomography (CT) scan for CT attenuation correction (CTAC) or CT coronary artery calcium score (CAC) using our multi-slice CT modality in positron emission tomography/computed tomography (PET/CT). The inclusion criteria for this study comprised patients aged 18 years and older, those referred for MPI between January 2024 and June 2024, and patients with available and complete imaging data in the database. Exclusion criteria included patients aged 17 years and younger, those with missing or incomplete imaging records, patients who experienced protocol deviations that did not align with the routine MPI workflow during the study period, and those whose scans had technical failures such as corrupted imaging files, acquisition errors, or unacceptable image quality as determined by physicians.

Myocardial perfusion imaging (MPI) protocol

The MPI protocol was adapted from American Society of Nuclear Cardiology (ASNC) guideline17. It includes the indications, contraindications, patient preparation, selection and dosage of stressing agents as well as SPECT imaging schedule for conducting MPI studies.

SPECT modalities

A cardiac-dedicated CZT gamma camera (Discovery NM 530c; GE Healthcare, Milwaukee, WI, USA) and cardiac-dedicated conventional gamma camera (Ventri; GE Healthcare, Milwaukee, WI, USA) were used to collect SPECT data, designated as GC-1 and GC-2, respectively. The CT modality was not available on both systems; hence, the CT portion was acquired using a PET/CT scanner (GE Discovery MI Digital Ready; GE Healthcare, Milwaukee, WI, USA) that incorporated with 128-CT slices system. Table 1 shows the general properties of SPECT modalities incorporated in our MPI practice.

Table 1 General specifications of GC-1 and GC-2.

Activity prescription technique

All patients received Tc-99 m tetrofosmin (Myoview; GE Healthcare, USA), with activity prescribed using a weight-based approach. The prescribed activity varied based on the imaging system (GC-1 or GC-2) and the selected MPI protocol, whether a one-day or two-day protocol. Details of protocol selection and activity prescription for both systems are presented in Table 2.

Table 2 Activity prescription of Tc-99 m Tetrofosmin for one-day and two-day MPI protocol.

SPECT acquisition

The choice of a one-day or two-day MPI protocol was primarily based on the subject’s body mass index (BMI). Patients with a BMI of 35 or higher generally underwent the two-day protocol to minimize radiation exposure, as higher activity were required due to increased body weight. In contrast, those with a BMI of 35 or lower were typically assigned to the one-day protocol. However, in certain cases, additional factors such as patient logistics were also considered when determining the appropriate protocol. Similarly, the choice of imaging system, either GC-1 or GC-2, was also influenced by BMI. Subjects with a BMI under 35 were primarily assigned to GC-1, while those with a BMI of 35 or higher were generally assigned to GC-2 to ensure consistency in image quality18. However, some subjects with a BMI of 35 or higher may have been scanned using GC-1 if they could be accommodated comfortably within the gantry. Table 3 presents the SPECT acquisition protocol for GC-1 and GC-2.

Table 3 SPECT acquisition protocol for GC-1 and GC-2.

CT acquisition

As previously indicated, PET/CT was used as an external modality for CT image acquisition. Depending on subject’s coronary artery disease status, they may have undergone CT attenuation correction (CTAC) alone using CTAC protocol or in conjunction with CT coronary artery calcium score (CAC) using the CAC protocol. The CAC protocol was performed for all of subjects except those who had previous of coronary artery bypass grafts (CABG) surgery or those who had undergone triple-vessel percutaneous coronary intervention (PCI) with stents. Table 4 summarizes the details of both CT acquisition protocols.

Table 4 CT acquisition protocol for CTAC and CAC.

SPECT image reconstruction

SPECT image reconstruction parameters were taken from routine protocol. Gated images in GC-1 were reconstructed using the maximum likelihood-expectation maximization (MLEM) method, adapted to the 3D geometry of the CZT camera, with CT attenuation correction. Conversely, gated images from GC-2 were reconstructed using the resolution recovery conventional ordered subset expectation maximization (OSEM) method, also with CT attenuation correction. The choice of reconstruction parameters was based on recommendations from the local application specialist in consensus with reporting physician. Image reconstruction was performed using a dedicated workstation, Xeleris version 4.1 (GE Healthcare, Milwaukee, WI, USA) generating standard axes views (short axis, vertical long axis, horizontal long axis) and polar maps of the left ventricle for clinician’s review. Table 5 presents the details on the SPECT image reconstruction protocols.

Table 5 Image reconstruction protocol in GC-1 and GC-2.

Data analysis

Subject demographics such as age, sex, height and weight, were recorded from each subject’s image files in the Picture Archiving and Communication System (PACS) Synapse version 5.7.200 (Fujifilm Medical Systems, USA). BMI was calculated as weight (kilograms) divided by the square of the height (metre2). For SPECT data collection, the scanner choice and administered activity for both stress and rest phases were recorded. For CT data, tube voltage, tube current, scan length, CT dose index volume (CTDIvol) and dose length product (DLP) were noted. Data analysis was conducted using IBM SPSS Statistics (Version 30.0; IBM Corp., Armonk, NY, USA) to calculate mean, standard deviation, 50th percentile, and 75th percentile values for both SPECT and CT data. Additionally, CT parameters such as DLP and CTDIvol were tabulated and presented in terms of minimum, maximum, mean ± standard deviation, 50th percentile and 75th percentile values. As previously described, the DRL was assigned to 75th percentile values, while the AD was attributed to 50th percentile values from the tabulated data for both SPECT and CT portions, respectively.

Effective dose (ED) estimation

The most common parameter to assess the radiation burden across different imaging modalities, scanner and protocol is the effective dose (ED), measured in millisieverts (mSv). ED characterizes whole-body exposure from non-uniform radiation exposure as a weighted average of organ absorbed doses. In this study, the effective dose calculation consists of two distinct estimations. The SPECT scan provides information on internal radiation exposure, requiring the computation of internal dosimetry, while CT scan assess external radiation exposure, necessitating the calculation of external beam dosimetry.

Effective dose SPECT (EDSPECT)

The ED estimation in SPECT, EDSPECT was calculated using Eq. 1 based on effective dose conversion coefficient per unit administered activity developed by Anderson. These coefficients, were derived and simulated using ICRP/ICRU biokinetic data, computational voxel reference phantoms, and tissue weighting factors from ICRP publications 103 specifically for Tc-99 m tetrofosmin stress and rest subjects. The corresponding values were 0.00515 mSv/MBq (male, stress), 0.00557 mSv/MBq (male, rest), 0.0062 mSv/MBq (female, stress), 0.00672 mSv/MBq (female, rest), respectively19,20.

$${ED}_{SPECT}\left(mSv\right)=Activity\:Administered\:\times\:\frac{Effective\:Dose\:Coefficient}{Unit\:Adminstered\:Activity}$$
(1)

Effective dose CT (EDCT)

The simplest method and most commonly used method to estimate the ED for CT scans (EDCT) is by multiplying the dose-length-product (DLP), a CT radiation parameter, by an effective dose conversion coefficient, known as a k-factor. The DLP, which is specific to CT, is reported on the scanner console after each acquisition and reflects for both the radiation exposure intensity in milliGray (mGy), and the craniocaudal length irradiated, measured in centimetre (cm). In this study, a cardiac-specific k-factor of 0.026 mSv·mGy− 1cm− 1 as proposed by Trattner et al., was applied to determine the EDCT for both CT attenuation correction (CTAC) protocol and coronary artery calcium score (CAC) protocol, using Eq. 221.

$${ED}_{CT}\:\left(mSv\right)=Dose\:Length\:Product\:\times\:{k-factor}$$
(2)

Cumulated effective dose (EDCUMULATED)

The cumulated effective dose (EDCUMULATED) associated with the combination of EDSPECT and EDCT, resulting from the hybrid imaging technique used in our MPI protocol was calculated using Eq. 3 in the unit of millisieverts (mSv). Since EDCUMULATED represents the combined dose from both SPECT and CT, the relative percentage contribution from each modality was determined to observe their respective contributions to the total radiation exposure.

$${ED}_{CUMULATED}\left(mSv\right)={ED}_{SPECT}+{ED}_{CT}$$
(3)

Results

Demographics

A total of 737 subjects were enrolled in this six-month observational study, with 676 subjects assigned to the one-day protocol and 61 subjects to the two-day protocol, as presented in Table 6. Of those assigned to the one-day protocol, 636 subjects underwent imaging with GC-1, while 40 subjects were scanned with GC-2. Conversely, of the 61 subjects in the two-day protocol, 18 were scanned in GC-1 and 43 using GC-2. The one-day protocol group included 481 male (71.2%) and 195 females (28.8%) with a mean age of 66.0 ± 10.2 years, a mean weight of 70.7 ± 12.2 kg and a mean BMI of 26.5 ± 3.9 kg/m2. Meanwhile, the two-day protocol group comprised 35 males (57.4%) and 26 females (42.6%), with mean values of 62.0 ± 12.4 years for age, 92.7 ± 16.0 kg for weight and 35.6 ± 5.7 kg/m2 for BMI. These findings clearly indicate that the one-day protocol had a significantly larger sample size than the two-day protocol.

Table 6 Subjects demographics across the one-day and the two-day MPI protocol.

SPECT data

The administered activity between the one-day and two-day protocols was examined without distinguishing between stress-rest or rest-stress sequence, as presented in Table 7. For the one-day protocol, the mean, AD, and DRL for the first injection were 250.02 ± 90.95 MBq, 240.50 MBq, and 263.63 MBq, respectively while for the second injection, these values were 696.98 ± 153.26 MBq, 691.90 MBq, and 777.93 MBq. In contrast, for the two-day protocol, the mean, AD, and DRL for the first injection were 985.93 ± 359.15 MBq, 1106.30 MBq, and 1239.50 MBq, and for the second injection, 942.21 ± 362.11 MBq, 1073.00 MBq, and 1221.00 MBq, respectively. These results indicate that the administered activity in the two-day protocol was higher than in the one-day protocol for both injections (p < 0.001).

Table 7 Summary of administered activity forTc-99 m Tetrofosmin in SPECT.

CT data

The summary of CT result is presented in Table 8. Two distinct CT protocols were utilized; CT attenuation correction (CTAC) and coronary artery calcium score (CAC) with the choice of protocol depending on the subject’s condition. Among the 737 subjects, 621 underwent CAC protocol, while the remaining subjects underwent the CTAC protocol. For CTAC protocol, the mean, AD, and DRL for the DLP values were 58.5 ± 30.1 mGy.cm, 53.1 mGy.cm, and 78.6 mGy.cm, respectively. The corresponding values for the CTDIvol were 3.1 ± 1.6 mGy, 2.8 mGy, and 4.3 mGy, respectively. In contrast, for the CAC protocol, the DLP had mean, AD and DRL values of 72.3 ± 10.3 mGy.cm, 67.8 mGy.cm, and 77.5 mGy.cm, respectively, while the CTDIvol values were 4.9 ± 0.2 mGy, 4.8 mGy, and 4.8 mGy, respectively. The mean scan lengths were 156.0 ± 21.0 mm for CTAC and 145.0 ± 21.1 mm for CAC. Overall, the values for CAC protocol were significantly higher across all parameters compared to those for CTAC.

Table 8 Summary of CT dose analysis by protocol.

Comparison with other similar findings

A comparison of both SPECT and CT data was made with findings from other studies. As recommended by International Commission on Radiological Protection (ICRP) Publication 135, the comparison primarily prioritized the 75th percentile, or known as DRL, while the 50th percentile, or known as AD, was used only when the DRL from the referenced study was unavailable15.

SPECT data

The comparison of SPECT data with other studies is shown in Table 9. It can be observed that the lowest DRL and AD for both the first and second injections in one-day tetrofosmin protocol were recorded in our study compared to similar studies. In contrast, for the two-day protocol, the DRL value of administered activity for both injections were higher than in the listed studies, except for the study by Alkhybari et al., whose findings were reported using AD, in which both of our injection values were lower than theirs22. As previously indicated, the majority of subjects in the two-day protocol were assigned to GC-2, which required a higher prescribed activity, thus leading in greater radiation dose to patients.

Table 9 Comparison of MPI practice against other published MPI studies in terms of Tc-99 m administered activities in first and second injection.

CT data

A comparison of CTDIvol and DLP in our MPI practice with similar studies are detailed in Table 10. As mentioned earlier, all CT protocols in the MPI study were performed using our PET/CT. Except for data reported by Abe et al. in PET/CT modality, our DRL for CTDIvol in CTAC was higher than those in the other listed studies25. Since Alkhybari et al. reported their findings using AD, it can be observed that our AD value of CTDIvol is lower than their reported value. A similar pattern was observed for CTDIvol value in the CAC protocol. In contrast, for DLP in CTAC, our DRL value was higher than in all listed studies except for those by Abe et al. and Rinscheid et al.25,29. Again, Since Alkhybari et al. reported their finding using AD, our AD value for DLP is lower than theirs. A similar trend was also observed for DLP in the CAC protocol.

Table 10 Comparison of CT in MPI against other published MPI studies in terms of CTDIvol and DLP.

Effective dose (ED) data

Table 11 shows the estimated effective dose (ED) from the MPI procedure incorporating contribution from both CT and SPECT imaging modalities. In the one-day and two-day protocol, the mean EDSPECT were 5.42 ± 1.33 mSv and 11.10 ± 3.96 mSv, respectively. The mean EDCT for CTAC was 1.42 ± 0.78 mSv in the one-day protocol and 2.26 ± 0.87 mSv in the two-day protocol, respectively. In contrast, the mean EDCT of CAC were 1.87 ± 0.27 mSv in the one-day protocol and 1.94 ± 0.31 mSv in the two-day protocol.

Table 11 Distribution of cumulated effective dose for across MPI protocols.

Figure 1 illustrates the average EDCUMULATED for each protocol category. The mean EDCUMULATED was 6.69 ± 1.76 mSv for the one-day/CTAC, 7.32 ± 1.37 mSv for the one-day/CAC, and 10.13 ± 5.31 mSv for the two-day/CTAC, and 13.99 ± 3.24 mSv for the two-day/CAC, respectively. The highest mean EDCUMULATED was observed in the two-day/CAC protocol, while the lowest was observed in the one-day/CTAC protocol.

Fig. 1
figure 1

Distribution of cumulated effective dose for across MPI protocols.

The relative contribution of ED from SPECT and CT was analysed and is further described in Table 12. On average, SPECT accounted for 78.3% of the total ED, while CT contributed 21.7%. This finding indicates that the administered activity of Tc-99 m tetrofosmin was the primary contributor to the total ED in MPI practice.

Table 12 Relative contribution of SPECT and CT in MPI protocol.

Discussion

The mean administered activity in the one-day protocol MPI is lower than the two-day protocol in both injections. In addition to having lower body weight, the majority of subjects in the one-day protocol were scanned using GC-1, a cardiac-dedicated SPECT featuring CZT technology34. This system requires lower administered activity of Tc-99 m tetrofosmin while providing shorter scan time and superior image quality.

Next, both the mean CTDIvol and DLP values in the CAC protocol are higher values compared to CTAC. This is because the CAC protocol utilizes CT axial acquisition, gated scan mode with fixed scan parameters resulting in higher tube current and voltage, which generally leads to higher radiation dose. The CAC protocol is a favorable element in clinical practice as it provides valuable information associated to perfusion findings, including tissue characterization and calcium score determination in the coronary arteries11,12. Furthermore, the DRL and AD values of CTDIvol in CAC protocol are comparable due to the fixed nature of the protocol parameters, which allow for nearly consistent CTDIvol normalisation using a 32 cm body phantom, despite variations in the subjects’ actual body proportions. Although not analyzed in the study, the size-specific dose estimate (SSDE) method, as proposed by the American Association of Physicists in Medicine (AAPM) Report 204, may serve as an alternative approach to determine actual CTDIvol. The SSDE calculation is based on conversion factors estimated from the subjects’ actual body size offering a more individualized dose assessment.

Comparison with other similar findings

SPECT

There are several possible causes for variations in DRL and AD for MPI studies in our data compared to the others. First, the variations of body habitus and weight across regions can impact radiation doses. Heavier patients may require higher doses to obtain acceptable image quality22. Apart from that, variations in equipment might also play a role. It has been widely reported that CZT based technology is superior to conventional SPECT system35,36. With CZT, a low radiation dose is required to produce adequate image quality. When compared to a center without CZT based technology, a center with it may report lower DRL and AD values. Regional practice patterns may also have an impact, as seen by the distinctly varied practices observed across different global locations, which suggests, a lack of standardization in MPI practice24. Nonetheless, several well recognized guidelines could serve as foundation for practice adaptation before being modified to suit local preferences17,37. In short, differences in patient body habitus, imaging protocols, equipment, and regional practice patterns all likely contribute to the variations seen in MPI DRLs worldwide. Standardizing practices globally could help optimize radiation doses for these common nuclear cardiology procedures.

CT

The main factors contributing to variations in CT doses in MPI studies compared to other similar studies can stem from several factors such as local parameters technical preferences, scanner technology differences, scan indications, radiographer practices and patient characteristics38,39. Therefore, standardizing scanning protocols and educate the operators on the importance of consistent scan lengths are crucial steps to minimize dose variations. Understanding these contributing factors can help in optimizing protocols and reducing unnecessary radiation exposure to patients, ultimately improving the quality and safety of MPI studies.

Effective dose (ED)

American Society of Nuclear Cardiology (ASNC) has stated that ED of 9 mSv or below is considered as the best MPI study practice. In our data, radiation doses of < 9 mSv were achieved in the one-day protocol but not in the two-day protocol. Based on this finding, there is a need for reduction of radiation exposure in the practice. Identifying on which modality carries the highest radiation burden is crucial, as this information can guide the development of a radiation dose optimization strategy. The result of this study showed that SPECT contributes a higher radiation dose than CT. Consequently, optimization efforts for SPECT should begin with a review of the activity prescription protocol, with particular attention to the two-day protocol. Given that most patients undergoing the two-day protocol are assigned to GC-2, optimization should focus on refining practices associated with GC-2. There are several ways to reduce radiation exposure in GC-2 protocol. Firstly, to be more proactive in performing stress-only imaging procedure. Alertness to the normal stress images is needed to avoid the subsequent rest imaging, whenever applicable. By doing this, up to 80% of radiation dose reduction could be achieved24,40,41.

Furthermore, a camera-based reduction strategy could also help to reduce radiation dose. This includes incorporating alternative settings in iterative reconstruction, resolution recovery, and noise modulation, that cope with lower cardiac SPECT counting statistics and thereby provide acceptable image quality despite reduced SPECT acquisition time and/or reduced injected radiopharmaceutical activity. By doing this, activity prescriptions could be set lower, even for heavier patient42,43,44. However, any changes to image reconstruction parameters should be made in close collaboration with the reporting clinician to ensure that the adjustments do not significantly alter the image quality from the perspective of the image interpreter. Another recommendation is to consider for positron emission tomography (PET) MPI, as it could help to reduce radiation dose. However, its feasibility may be limited due to the high cost of the procedure and the availability of PET radiopharmaceuticals41,45,46.

This study has several limitations. Since the study was conducted at a single tertiary cardiac center, the results may not be fully representative of other institutions with different imaging equipment, protocols, or patient demographics. Future studies should adopt a multi-centre approach involving diverse institutions with varying imaging protocols, equipment types, and patient demographics. This would enhance the robustness of the data, ensuring the established DRLs and ADs are more representative of national or regional practices. Next, as the study only included adult subjects, the established DRLs and ADs may not apply to younger populations requiring distinct imaging protocols. Future research should focus on conducting a separate study specifically targeting paediatric and younger adult populations to establish appropriate DRLs and ADs for these groups.

Lastly, while this study proposed optimization strategies, no active intervention was conducted to assess their clinical impact. Future efforts should focus on implementing these strategies and evaluating their effectiveness in clinical practice. This is particularly important given the higher ED observed in patients scanned using GC-2, emphasizing the need for targeted interventions to reduce radiation exposure while maintaining diagnostic accuracy. Recognizing this, we have acknowledged it as a study limitation to highlight the need for future research exploring active intervention strategies for dose optimization and their potential impact on the generalizability of the established DRLs.

Conclusion

Local DRL and AD data for MPI procedures have been successfully established, along with the calculation of ED from the MPI study to assess the variations in radiation doses received from each component. These data serve as supplementary information to support professional judgments that weigh the benefits and risks of ionizing radiation in medical imaging, rather than as a constraint on the appropriate use of medical radiation. To ensure patient safety, applying the as low as reasonably achievable (ALARA) principle consistently in medical radiation practices is needed. Consequently, this ensures that patient safety remains a top priority while still achieving adequate diagnostic outcomes. Establishing local DRLs and the calculation of ED are valuable tools for guiding radiological practices, but they should complement, not replace, clinical judgment. By adhering to the ALARA principle, healthcare professionals can effectively balance the need for diagnostic accuracy with the imperative to minimize radiation exposure, ultimately contributing to safer and more effective medical imaging practices.