Table 4 CT acquisition protocol for CTAC and CAC.

From: Diagnostic reference level, achievable dose, and effective dose estimation in adult hybrid SPECT myocardial perfusion imaging in Institut Jantung Negara

System

GE discovery MI digital ready with 128 CT slices

Acquisitions

CTAC

CAC

Mode

Low-Dose, Non-Gated

Prospective-ECG Gated

Scan type

Spiral

Axial (Cine Mode)

Tube voltage

120 kVp

120 kVp

Tube current

30–100 mA

215 mA

Rotation time

0.8 s

0.35 s

Rotation length

Full

Segment

Detector coverage

40 mm

20 mm

Image thickness in recon 1 (Primary Recon)

Helical thickness—5.0 mm

Axial thickness—2.5 mm

Pitch—0.984

Image per rotation—8 images

Speed—38.37 mm/rot

Cine between images—0.05 s

Reconstructed image

CT Cardiac and CTAC

CT calcium score, CT Cardiac and CTAC

Reconstructed DFOV

Recon 1—50 cm

Recon 1—25 cm

Recon 2—50 cm

Recon 2—41.9 cm

Recon 3—41.9 cm

Reconstructed image slice thickness

Recon 1—CTAC – 5.00 mm

Recon 1—CAC – 2.5 mm

Recon 2—CTAC – 5.00 mm

Recon 2—CT Cardiac – 1.25 mm

Recon 3—CT Cardiac – 1.25 mm

Matrix size

512 × 512 mm

Anatomy region

Carina to apex of the heart