Table 1 Profiles of the portable infusion pumps evaluated in this study.
From: Accuracy of infusion flow rates and bolus doses for portable infusion pump
Pump | Reservoir volume (mL) | Adjustable flow rate (mL/h) | Power source | Calibration temperature (°C) | Flow rate accuracy* | Weight (g) |
|---|---|---|---|---|---|---|
COOPDECH Amy PCA | 100 | 0.1–30.0 | Electronic (micropump) | 25 | ± 6% | 140 |
CADD-Solis | 100 | 0–100 | Electronic (squeezing) | 2–40 | ± 6% | 595 |
COOPDECH Balloonjector | 200 | 2, 4, 6 | Elastomeric (balloon) | 23 | ± 10% | 153 |