Table 2 Average measured flow rate as a percentage of the set infusion rate (4 mL/h) during the dosing period (%).
From: Accuracy of infusion flow rates and bolus doses for portable infusion pump
Pump | Control | Epidural catheter | Motion | Warming | P value |
|---|---|---|---|---|---|
COOPDECH Amy PCA | 95.2 (1.9) | 96.0 (2.7) | 91.8 (3.8) | 97.0 (3.2) | 0.231 |
CADD-Solis | 98.1 (1.6) | 93.1 (0.5)*1 | 102.4 (0.7)*2 | 99.3 (1.7) | < 0.001 |
COOPDECH Balloonjector | 98.2 (2.5) | 92.3 (3.0) | 91.4 (1.0) | 117.5 (0.7)*3 | < 0.001 |
Rakurakufuser | 78.9 (3.6) | 78.8 (2.3) | 78.5 (3.3) | 99.3 (3.9)*4 | < 0.001 |