Table 3 Percentage of dosing time within 100% ± 15% of set flow rate (4 mL/h) during the dosing period (%).
From: Accuracy of infusion flow rates and bolus doses for portable infusion pump
Pump | Control | Epidural catheter | Motion | Warming | P value |
|---|---|---|---|---|---|
COOPDECH Amy PCA | 93.2 (8.4) | 88.1 (17.2) | 89.6 (13.6) | 94.8 (2.8) | 0.887 |
CADD-Solis | 96.3 (1.4) | 94.0 (1.3) | 95.1 (1.3) | 93.4 (4.3) | 0.521 |
COOPDECH Balloonjector | 65.5 (8.9) | 68.3 (10.5) | 66.1 (10.1) | 43.3 (4.7) | 0.027 |
Rakurakufuser | 31.0 (12.2) | 17.4 (5.8) | 18.8 (14.9) | 89.1 (0.5) * | < 0.001 |