Table 4 Measured volumes of bolus doses (3 mL).

From: Accuracy of infusion flow rates and bolus doses for portable infusion pump

Pump

Control

Epidural catheter

Motion

Warming

P value

COOPDECH Amy PCA

3.0 (0.0)

3.0 (0.1)

3.1 (0.1)

3.0 (0.0)

0.499

CADD-Solis

3.0 (0.1)

2.7 (0.1)*

3.1 (0.0)

3.1 (0.1)

 < 0.001

COOPDECH Balloonjector

3.0 (0.0)

3.0 (0.1)

2.9 (0.1)

3.0 (0.1)

0.465

Rakurakufuser

3.0 (0.1)

3.0 (0.1)

3.1 (0.1)

3.1 (0.2)

0.445

  1. Data are presented as mean (standard deviation).