Table 4 Measured volumes of bolus doses (3 mL).
From: Accuracy of infusion flow rates and bolus doses for portable infusion pump
Pump | Control | Epidural catheter | Motion | Warming | P value |
|---|---|---|---|---|---|
COOPDECH Amy PCA | 3.0 (0.0) | 3.0 (0.1) | 3.1 (0.1) | 3.0 (0.0) | 0.499 |
CADD-Solis | 3.0 (0.1) | 2.7 (0.1)* | 3.1 (0.0) | 3.1 (0.1) | < 0.001 |
COOPDECH Balloonjector | 3.0 (0.0) | 3.0 (0.1) | 2.9 (0.1) | 3.0 (0.1) | 0.465 |
Rakurakufuser | 3.0 (0.1) | 3.0 (0.1) | 3.1 (0.1) | 3.1 (0.2) | 0.445 |