Table 4 Parameters to analyze the link budget of the proposed WPT system.

From: Development of an efficient mid-field wireless power transmission system for biotelemetric IoT medical devices

\({{\textbf {T}}}_{{\textbf {x}}}\)-antenna characteristics

\({{\textbf {R}}}_{{\textbf {x}}}\)-antenna characteristics

Frequency

1.5 GHz

Frequency

1.5 GHz

Transmitted power (\({\text {P}}_{\text {t}}\))

30 dBm

Receiver gain (\({\text {G}}_{\text {r}}\))

-26 dBi

\({\text {T}}_{\text {x}}\) antenna gain (\({\text {G}}_{\text {t}}\))

1 dBi

Feeding loss

0 dBm

Feeding loss (\({\text {L}}_{\text {f}}\))

0 dBm

Radiation efficiency

0.95 %

EIRP (\({\text {P}}_{\text {t}}\) + \({\text {G}}_{\text {t}}\) - \({\text {L}}_{\text {f}}\))

31 dBm

Polarization loss

0 dBm

WPT-link budget characteristics

Power received (\({{\textbf {P}}}_{{\textbf {r}}}\))

Distance (cm)

5

\({\text {P}}_{\text {r}}\) at 2 cm

13.5 dBm

Cable loss (dBm)

1

\({\text {P}}_{\text {r}}\) at 3 cm

12.5 dBm

Path loss (dBm)

1,2,3,4

\({\text {P}}_{\text {r}}\) at 4 cm

9.25 dBm

Mismatch loss (dBm)

0

\({\text {P}}_{\text {r}}\) at 5 cm

8.28 dBm