Table 4 Multi-zone high-temperature furnace supplementary test point temperature field data sheet.

From: Development and testing of a high-precision static calibration device for aviation thin-film thermocouples

Axial coordinate

200℃

400℃

600℃

800℃

1000℃

1200℃

-50

1.306

1.271

1.352

0.587

0.287

-0.242

-40

0.835

0.895

0.745

0.238

0.208

-0.128

-30

0.336

0.423

0.584

0.151

-0.004

-0.137

-20

-0.003

0.064

0.230

0.078

-0.079

-0.077

-10

-0.105

-0.097

0.000

0.065

-0.038

-0.014

0

0.000

0.000

0.000

0.000

0.000

0.000

10

0.278

0.195

0.179

-0.061

0.093

0.022

20

0.593

0.479

0.418

-0.112

0.250

0.129

30

0.838

0.645

0.482

-0.203

0.505

0.297

40

0.676

0.556

-0.901

-0.490

0.665

0.437

50

-1.350

-0.979

-1.580

-1.462

-0.073

0.247