Fig. 3: SMELL-S results. | Communications Engineering

Fig. 3: SMELL-S results.

From: Multi-channel portable odor delivery device for self-administered and rapid smell testing

Fig. 3

a Odor intensity measured using the photoionization detector (PID) for the SMELL-S odor stimuli. Mean and SD were calculated from a set of n = 10 repeat measurements. b An example of a subject’s performance during the SMELL-S test, for a patient with smell loss treated at the Geneva University Hospital (orange line) and a subject with an intact sense of smell (yellow line). The task becomes more and more difficult as the level approaches 10. A reversal occurs when the direction in which the concentration is changed reverses (blue dots). The score is defined as the mean of the last four reversals (gray dotted lines). c Violin plots showing the durations for the Sniffin' Sticks and SMELL-S threshold tests. Lines indicate the median, 25% and 75% quartiles and extrema.