Table 1 The simulated and measured FWHM along the lateral and axial direction of holograms with different focal length

From: Acoustic hologram-enabled simultaneous multi-target blood-brain barrier opening (AH-SiMBO)

Hologram

Hologram 1

Hologram 2

Hologram 3

Focal Length

10 mm

20 mm

30 mm

 

Sim (mm)

Mes (mm)

Diff

Sim (mm)

Mes (mm)

Diff

Sim (mm)

Mes (mm)

Diff

Lateral FWHM

0.90

0.78

13.33%

1.30

1.08

16.92%

1.70

1.37

19.41%

Axial FWHM

3.40

2.72

20.00%

6.00

5.88

2.00%

13.10

11.68

10.84%

Actual focal length

10.7

10.8

0.93%

20.9

21.0

0.48%

30

30.20

0.67%

  1. Diff difference.