Table 3 The simulated and measured FWHM along the lateral and axial direction of holograms with different rotation angles
From: Acoustic hologram-enabled simultaneous multi-target blood-brain barrier opening (AH-SiMBO)
Angle | 3° | 6° | 9° | 15° | 30° | ||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
Sim (mm) | Mes (mm) | Sim (mm) | Mes (mm) | Sim (mm) | Mes (mm) | Sim (mm) | Mes (mm) | Sim (mm) | Mes (mm) | Average diff | |
Lateral FWHM | 1.30 | 1.09 | 1.30 | 1.09 | 1.30 | 1.11 | 1.50 | 1.15 | 1.50 | 1.24 | 17.52% |
Axial FWHM | 5.40 | 6.03 | 5.40 | 6.07 | 5.20 | 5.58 | 4.90 | 5.34 | 4.20 | 4.90 | 11.41% |