Table 3 The simulated and measured FWHM along the lateral and axial direction of holograms with different rotation angles

From: Acoustic hologram-enabled simultaneous multi-target blood-brain barrier opening (AH-SiMBO)

Angle

15°

30°

 
 

Sim (mm)

Mes (mm)

Sim (mm)

Mes (mm)

Sim (mm)

Mes (mm)

Sim (mm)

Mes (mm)

Sim (mm)

Mes (mm)

Average diff

Lateral FWHM

1.30

1.09

1.30

1.09

1.30

1.11

1.50

1.15

1.50

1.24

17.52%

Axial FWHM

5.40

6.03

5.40

6.07

5.20

5.58

4.90

5.34

4.20

4.90

11.41%

  1. Diff difference.