Table 2 Recovery study with blind spiked samples with the multiplexed plasmonic biosensor

From: Disposable Point of Care multiplexed plasmonic biosensor for rapid and specific identification of respiratory viruses

Samples

Spiked NP

Measured NP

Recovery (%)

Virus

[NP] (ng/mL)

CH1 Anti-IBV

CH2 Anti-RSV

CH3 Anti-IAV

CH4 Anti-COV

[NP] (ng/mL)

ΔλCH1 (nm)

ΔλCH2 (nm)

ΔλCH3 (nm)

ΔλCH4 (nm)

S1

Influenza B

4000

(+) 2.90

(-) 0.02

(-) 0.00

(-) 0.04

3729.7

93.2

S2

Influenza B

1000

(+) 1.01

(-) 0.03

(-) 0.02

(-) 0.00

1069.4

106.9

S3

Influenza A

500

(-) 0.03

(-) 0.01

(+) 0.23

(-) 0.01

791.4

158.3

S4

SARS-CoV-2

1800

(-) 0.03

(-) 0.04

(-) 0.04

(+) 2.69

2689.8

149.9

S5

RSV B

10

(-) 0.02

(-) 0.00

(-) 0.03

(-) 0.02

< LOD

-

S6

Influenza A

3500

(-) 0.03

(-) 0.00

(+) 0.70

(-) 0.03

3312.4

94.6

S7

Influenza A

1200

(-) 0.02

(-) 0.02

(+) 0.40

(-) 0.02

1700.9

141.7

S8

Influenza B

100

(+) 0.13

(-) 0.02

(-) 0.03

(-) 0.04

128.7

128.7

S9

SARS-CoV-2

50

(-) 0.03

(-) 0.03

(-) 0.03

(+) 0.15

46.62

93.2

S10

Influenza B

1500

(+) 1.25

(-) 0.00

(-) 0.00

(+) 1.69

1353.2

90.2

SARS-CoV-2

800

866.9

108.3

S11

Influenza A

2000

(-) 0.04

(-) 0.02

(+) 0.41

(+) 0.76

1700.9

85.1

SARS-CoV-2

400

269.1

67.2

  1. *Sample was considered positive (+) or negative (-) for a given virus for signals above the set threshold (Δλ < 0.059 nm) or below (Δλ < 0.058 nm). Samples in between those values would be considered “indetermined”. Determination of the cut-off values is described in the Experimental section.