Introduction

Osteoarthritis (OA), the most prevalent progressive musculoskeletal condition, has garnered substantial research interest1. Single-compartment lesions are common, representing 50% of cases, with isolated medial tibiofemoral OA comprising 27%2. Unicompartmental knee arthroplasty (UKA) provides several benefits over total knee arthroplasty (TKA) for treating isolated medial tibiofemoral OA, with more bone preserving, earlier recovery, and more natural knee sensation3,4,5.

Despite significant advancements in UKA surgical techniques and implant designs over the past decades, the higher postoperative revision rate for UKA (18.2% at 15 years) compared to TKA (6.2% at 15 years) remains a significant obstacle to its broader adoption6, which includes aseptic loosening, polyethylene (PE) wear, and the development of arthritis in uninvolved compartments7,8. Previous studies have identified abnormal contact stress on the articular surface due to prosthesis malalignment as a primary cause of revisions9,10,11. The tolerance for faults in UKA prostheses is lower than TKA because more anatomical structures are preserved9. While the limited surgical exposure in UKA benefits the patient, it also poses a greater technical challenge for accurate component placement, making it more prone to errors12. Prosthesis misalignment can result in excessive stress on the PE insert or healthy compartments’ surfaces, accelerating wear and leading to inflammatory responses that ultimately cause prosthesis loosening and cartilage degeneration13,14. Therefore, in the context of UKA, the alignment and positioning of the components are deemed essential for optimal clinical outcomes and longevity. Extensive research has documented the biomechanical changes and clinical results related to the tibial component’s positioning10,11,15,16. However, investigations into the biomechanical consequences of femoral component malalignment in UKA remain limited9.

To our knowledge, biomechanical studies on fixed-bearing prostheses are still insufficient, despite their higher wear rates17 and the increased precision required in prosthesis positioning18 compared to mobile-bearing prostheses. Notably, the influence of femoral implant misalignment on stress distribution across major knee ligaments, such as the medial and lateral collateral ligaments (MCL and LCL).

This research is designed to elucidate the biomechanical effects of femoral prosthesis misalignment as well as its impact on the main ligaments using finite element analysis (FEA). Various UKA models with differing femoral prosthesis rotations on the sagittal plane and translations on the coronal plane will be analyzed.

Materials and methods

Native knee model

Imaging data were collected from a volunteer (23 years, 170 cm, and 60 kg), with no known musculoskeletal disorders. Informed consent for his participation and imaging procedures was gathered. Scanning of the left knee was conducted with a 1.5 Tesla magnetic resonance imaging (MRI) scanner and a 64-slice computed tomography (CT) scanner (Siemens, Germany). All human-involved procedures in this study were conducted in alignment with the ethical standards of the institutional and/or national research council, in addition to adhering to the 1964 Declaration of Helsinki and its later amendments or comparable ethical norms. All experimental protocols were approved by the Institutional Review Board at our hospital (the First Affiliated Hospital of Wannan Medical College).

CT scans were conducted with a 0.6 mm thick slice, and MRI scans with a 1.0 mm thick slice, both in the sagittal plane, covering 15 cm of both the proximal tibia and distal femur. CT and MRI images obtained were then imported into Mimics software (version 20, Materialize, Leuven, Belgium) to generate 3D representations. These models were subsequently exported in stereolithography (STL) format and further processed using GeomagicWrap software (version 12.0, Geomagic, North Carolina, USA). The solid model was assembled and constructed in SolidWorks (version 2022, Dassault Systèmes S.A., USA) (Fig. 1).

UKA knee model

Upon creating the native knee model, an appropriately sized fixed-bearing UKA prosthesis (Link-Sled®, Germany) was selected. This prosthesis was scanned using the KSCAN-Magic 3D scanner (Aishankin Design Services Ltd., China). The refined and smoothed solid model of the prosthesis was imported into Abaqus 6.14 (Dassault Systèmes Simulia, Inc., France) for meshing, with detailed specifications provided in Table 1. All materials in this FE model were characterized by linear-elastic isotropic behavior, with mechanical properties defined based on previous studies9,19 (Table 2). Boolean operations were employed to simulate the medial UKA of the knee, ensuring that the PE implant’s height aligned with the native bone anatomy through a sagittal cut. The tibial implant was designed to cover the osteotomy surface optimally, avoiding inversion, eversion, and axial rotation.

Table 1 The meshing of each anatomical structural unit.
Table 2 Material properties of different structures used in this study.

In line with prior research9,20, a posterior slope of 5° was established via a transverse cut to align with the patient’s native tibial slope. The femoral implant was positioned perpendicular to the tibial prosthesis in the coronal plane, without axial rotation, covering the center of the medial femoral condyle and the distal end of the posterior femoral condyle as comprehensively as possible. A 1 mm thick layer of bone cement (methyl methacrylate) was simulated between the osteotomy surface and tibial implant to complete the fixed-bearing UKA model in its standard position, defined as a linearly elastic isotropic material (Table 2; Fig. 1).

Deviated prostheses were modeled with 3 mm and 5 mm of medial and lateral translation, as well as 3° and 5°of internal and external rotation, using the midline of the femoral prosthesis as a reference, as depicted in Fig. 2. The interface between cartilage and bone within the model was treated as fully bonded. All contact surfaces in the fixed-bearing UKA model utilized a finite sliding frictionless hard contact algorithm without penetration9,21. Convergence within the model was defined by a relative change of greater than 5% between two consecutive mesh refinements. The average element size for the simulated cartilage and menisci was 0.8 mm 21.

Stress loading and model validation

A compressive load of 1000 N along the mechanical axis of the femur downwards was applied to the different models to examine stress distributions, with a focus on the upper surface of the polyethylene (PE) insert, the tibial cartilage surface in the lateral compartment, and the MCL and LCL. The boundary conditions of each model were kept constant during loading, and convergence was verified after loading9,19.

The FE knee model underwent validation through axial load testing and simulation of the anterior drawer experiment. For validation under axial load, the femur was immobilized in the flexion-extension direction, whereas the lower surfaces of the tibia and fibula were entirely constrained22. This experiment setup aimed to mimic a realistic, functional scenario to validate the FEA model by preventing unwanted rotation and movement while allowing for the application and measurement of axial forces. Cartilage-to-bone and ligament-to-bone attachments were modeled as bindings, with the anterior and posterior horns of the meniscus affixed to the tibial plateau. An axial load of 1150 N was applied along the mechanical axis23, and the resulting contact pressures were compared with data from previous research23. In the simulation of the anterior drawer experiment, tibiofibular flexion and femoral movement were restricted. An anterior load of 134 N, perpendicular to the coronal plane, was then applied at a reference point at the midpoint between the medial and lateral edges of the tibial plateau9,24. The observed tibial displacement was subsequently compared with findings from previous studies24.

Results

Validation of the UKA model

Under an axial load of 1150 N, the peak contact forces observed were 2.91 MPa for the medial meniscus and 1.49 MPa for the lateral meniscus. These values deviated by less than 3% from the contact pressures reported by Pena et al. 25. The maximum compressive stresses recorded in the medial and lateral menisci were 1.39 and 1.62 MPa, respectively, aligning closely with the values reported by Pena et al. (1.42 and 1.62 MPa)25. The simulated anterior drawer experiment revealed an anterior displacement of 4.92 mm at the tibial loading point, consistent with previous studies24.

Stress distribution on the PE insert

In the translation model, the stress on the PE insert’s upper surface was 38.19 MPa when the femoral prosthesis was positioned standardly. With lateral translations of 3 and 5 mm, the stress increased to 39.79 and 41.61 MPa, respectively. Conversely, medial translations resulted in decreased stress values, measuring 34.56 MPa for a 3 mm shift (a 9.51% reduction) and 32.94 MPa for a 5 mm shift (a 13.75% reduction).

Comparatively, the peak contact stresses on the medial compartment of PE insert exhibited notable increases at both varus and valgus angles of 5°, registering 42.52 and 42.95 MPa, representing increases of 11.34 and 12.46%, respectively. For varus and valgus angles of 3°, the maximum contact stresses (MCSs) were 39.74 and 38.40 MPa, which were 4.06 and 0.55% higher than the standard position. Details are illustrated in Figs. 3 and 4.

Stress distribution on the tibial cartilage of lateral compartment

In the translation model, the stress distribution on the tibial cartilage of the lateral compartment showed an opposite trend compared to the PE insert. Medial translations of 3 and 5 mm resulted in stress increases of 11.40 and 16.52%, respectively, with stress values of 7.82, 8.18, and 7.02 MPa for the standard position, 3 and 5 mm shifts. Conversely, lateral translations generally reduced stress on the cartilage surface, measuring 6.92, 6.59, and 7.02 MPa for 3, 5 mm translations, and the standard position, respectively.

In the varus/valgus model, the stress on the lateral compartment’s tibial cartilage decreased with valgus alignment, showing a reduction of 4.27% at 3° valgus and 8.26% at 5° valgus (with stress values of 6.72, 6.44, and 7.02 MPa for the standard position). This trend reversed with varus alignment, where a 3° varus increased stress by 2.28% and a 5° varus increased stress by 5.98% (with stress values of 7.18, 7.44, and 7.02 MPa for the standard position). Further details are provided in Fig. 4.

Stress distribution on MCL and LCL

In the translation model, the maximum principal stresses on MCL and LCL followed an opposite trend, generally decreasing and increasing with the lateral translation of the femoral implant. In the varus/valgus model, the maximum principal stresses on the MCL increased with the valgus alignment of the femoral prosthesis, rising by 11.34% at 3° and 11.94% at 5° (0.0335, 0.0373, and 0.0375 MPa for standard position, 3°, and 5°). Conversely, these stresses decreased with an increasing varus angle, showing reductions of 2.69% at 3° and 7.76% at 5°. Similarly, the maximum principal stresses in the LCL decreased by 5.00% at 3° valgus and 6.43% at 5° valgus (0.0140, 0.0133, and 0.0131 MPa for standard position, 3°, and 5°). These stresses increased with an increasing varus angle, with reductions of 0.71% at 3° and 5.00% at 5°. The detailed results are illustrated in Fig. 5.

Discussion

In treating single-compartmental knee osteoarthritis (KOA), UKA offers significant advantages over TKA, such as greater preservation of bone and cartilage tissues and improved postoperative knee function12. However, the limited wound exposure in UKA compared to TKA, particularly in obese patients, increases the risk of prosthesis misalignment, a primary factor in the need for revision surgery26,27. Reducing the revision rate of UKA prostheses remains a critical challenge, largely dependent on achieving proper alignment through accurate implant positioning and soft-tissue balancing12.

The appropriate fitting range for UKA prostheses, especially concerning the biomechanical impact of femoral implant positioning, remains uncertain. The biomechanical consequences of various placements of the femoral component within the knee joint are not yet fully comprehended. Jennings et al. 12 recommended positioning the femoral component centrally or slightly laterally on the medial femoral condyle to optimize tracking with the tibial component. Liu W and colleagues28 analyzed the effect of valgus/varus angles of the femoral components (Oxford mobile platform prosthesis) on mid-term outcomes post-UKA, revealing that an internal rotation angle between 0° and 6° could achieve superior mid-term results. Gulati et al. 29 conducted a follow-up study on 211 knees with medially implanted mobile-bearing UKAs using a minimally invasive technique over at least four years. They found that in 98% of cases, the femoral component was within 10° varus and 10° valgus, concluding that the Oxford UKA tolerates femoral malalignment of up to 10° due to its spherical femoral component. Unlike mobile-bearing prostheses, fixed-bearing prostheses cannot adjust for abnormal stress caused by misalignment, leading to increased edge loading and rapid wear, even with significant malalignment29. Consequently, fixed-bearing devices are more prone to deviation and less tolerant of fitting errors than mobile-bearing prostheses9. This study selected a fixed-bearing UKA prosthesis (Link-Sled®, Germany) to examine the mechanical effects of the femoral prosthesis on the internal structures of the knee joint.

Our findings align with previous studies9,21 that demonstrated the significant impact of the femoral component’s position on the contact stress experienced by the PE insert and the lateral compartment. In both varus and valgus fixed-bearing UKA models, there was an increase in contact stresses on the PE insert. However, an increase in contact stress within the lateral compartment was observed only in the varus model, supporting earlier findings30. This is attributed to the varying stiffness between the PE insert and the lateral compartment (elastic modulus: 685 MPa for the medial PE insert vs. 15 MPa for the lateral compartment)30. Additionally, a lateral deviation of 5 mm in the femoral prosthesis resulted in an 8.96% increase in contact stress on the PE insert, while the articular cartilage of the contralateral compartment exhibited a 6.13% reduction in contact stress relative to the standard positioning. Kang et al. 21 observed that lateral translations of 3 and 5 mm of the femoral component under gait loading conditions resulted in increases of 8.2 and 11.1% in MCS on the PE insert, respectively. These findings were further verified through static and dynamic modeling, demonstrating that femoral implant misalignment affects MCS on the PE insert and tibial cartilage of the lateral compartment.

This aligns with the earlier concept that the optimal implant position for UKA is to replace the original cartilage and replicate the natural joint line31. However, achieving a central femoral component placement in clinical settings is challenging due to factors like minimally invasive incisions and the absence of intraoperative proofreading instruments. Jaffry et al.32 compared the accuracy of UKA prosthesis positioning using patient-specific instrumentation (PSI) and robotic systems. Their results indicated that femoral implant positions and orientations were significantly more accurate with PSI, with mean errors of 2 mm and 4°, compared to 4 mm and 10° found in the conventional approach. Notably, no significant difference in accuracy was observed between the robotic and PSI methods.

A notable aspect of this research was the investigation of the effects of femoral prosthesis malalignment on stress distribution in MCL and LCL. The position of the femoral component was found to impact the maximum principal stresses on the MCL and LCL. Compared to the neutral position UKA FE model, the maximum principal stresses on the MCL increased in models with a valgus 5° or medial translation of 5 mm. Conversely, the stress distribution in the LCL increased less in the 5° valgus or 5 mm lateral translation models. This disparity may be due to the significant stiffness variability between the PE insert and lateral cartilage33. Thus, femoral prosthesis malalignment could alter stress in the MCL and LCL, potentially leading to stiffness or pain33.

This study does have some limitations. Firstly, while selecting a single volunteer for various deviation analyses eliminated heterogeneity, it also limited the representativeness of the results. KOA patients typically treated by UKA often have altered joint mechanics with different anatomical adaptations. The study will be expanded in the future to include an exploration of multiple prosthetic models representing different anatomical structures. Secondly, the study focused on static analysis and did not include gait cycle analysis under dynamic loading, which will be addressed in subsequent research. Thirdly, rotational activity of the knee was restricted under loading to simplify stress distribution testing. This aspect will be further evaluated in the future by in vitro mechanical testing. For instance, muscle tissue or dynamic stabilization effects will be incorporated into the model to assess the impact of malalignment more accurately on the PE and cartilage. Despite certain limitations, the insights garnered from this research provide valuable guidance for performing UKA.

Conclusions

The precise positioning of the femoral prosthesis in UKA, including varus, valgus angles, and translational shifts, significantly impacts the MCS on the PE insert and tibial cartilage, as well as the principal stresses in MCL and LCL. Surgeons must pay close attention to the alignment of the femoral prosthesis, particularly deviations exceeding 5°or 5 mm, as these misalignments can lead to increased contact stresses. To ensure the long-term survival of the prosthesis and enhance postoperative satisfaction, it is recommended that the femoral prosthesis be positioned centrally.

Fig. 1
figure 1

The finite element models of the intact knee and UKA in the standard position of the prosthesis.

Fig. 2
figure 2

Finite element models of UKA with different translations and varus/valgus of the femoral prosthesis. a1: 5 mm medial translations; b1: 3 mm medial translations; c: standard position; d1: 3 mm lateral translations; e1: 5 mm lateral translations; a2: 5° varus; b2: 3° varus; d2: 3° valgus; e2: 5° valgus.

Fig. 3
figure 3

Maximum contact stress distribution on PE insert with respect to a1: 5 mm medial translations; b1: 3 mm medial translations; c: standard position; d1: 3 mm lateral translations; e1: 5 mm lateral translations; a2: 5° varus; b2: 3° varus; d2: 3° valgus; e2: 5° valgus.

Fig. 4
figure 4

Maximum contact stress distributions on PE insert and lateral tibial cartilage with respect to medial and lateral translations, as well as varus and valgus. a1-a2: Upper surface of PE insert; b1-b2: surface of tibial cartilage in lateral compartment.

Fig. 5
figure 5

Maximum principle stress distributions on medial and lateral collateral ligaments (MCL and LCL) with respect to medial and lateral translations, as well as varus and valgus. c1-c2: MCL; d1-d2: LCL.

Structures

Units

Element node

Type of unit

Bone

62,268

78,886

C3D10M

Cartilage

74,369

124,483

C3D10M

Ligaments

52,739

85,725

C3D10M

Metal implants

54,820

96,009

C3D10M

PE insert

6834

8421

C3D8R

Meniscus

68,621

106,012

C3D10M

Table 2. Material properties of different structures used in this study.

Structures

Elastic modulus (MPa)

Poisson’s ratio

Cortical bone

17,000

0.3

Cancellous bone

350

0.25

Menisci

27.5

0.33

Cartilage

15

0.46

Cobalt-chromium-molybdenum alloy

210,000

0.29

Polyethylene (PE)

850

0.4

Methacrylate

1940

0.4