Introduction

Corneal refractive surgery is gradually becoming safer and more effective because of advancements in femtosecond and excimer laser technology. Before surgery, it has always been important to carefully determine the surgical indication1. Keratoconus (KC) is recognized as a contraindication for corneal refractive surgery and poses a significant risk to the patient’s vision. Therefore, it is crucial to conduct thorough screenings for patients who may exhibit signs of suspected keratoconus (refer to Fig. 1D). This condition is characterized by an abnormal steepening of the corneal topography, which may present as an asymmetric, truncated, or skewed-axis bowtie in the inferior focal area, while the fellow eye remains free of keratoconus2 (Fig. 1C). Subclinical KC (Fig. 1B) shares a similar topographic asymmetry with suspected KC, although the fellow eye has KC3 (Fig. 1A) and may progress to KC4. Thus, it is especially crucial to identify individuals who may have KC before they experience visual loss or exhibit a positive slit-lamp test result.

Fig. 1
Fig. 1
Full size image

Subclinical KC patient and Suspected KC patient. (A) Keratoconus(KC) eye, I-S Value = 1.73, BADD = 6.03; (B) Contralateral subclinical KC eye, I-S Value = 1.47, BADD = 1.93; (C) Normal eye, I-S Value = 1.08, BADD = 1.44; (D) Contralateral suspected KC eye, I-S Value = 1.61, BADD = 1.53.

The examination of corneal topography via various design principles is necessary to diagnose KC. The Pentacam can measure anterior and posterior corneal surface curvature and corneal thickness with accuracy via its unique Belin/Ambrósio system, whereas the Topometric/KC system shows significance in the diagnosis of KC5,6,7,8. According to recent reports9,10, KC can occur when corneal collagen fibers rupture in the early stages of the disease, and changes in corneal biomechanical characteristics occur earlier than changes in corneal morphological changes. Therefore, analyzing the combination of morphology and biomechanics via the Pentcam and Corvis ST to differentiate between KC and other conditions is a highly effective strategy11,12,13.

An increasing number of artificial intelligence-related models, such as decision trees, random forests, and machine learning, have been found to be applicable in predicting KC14,15,16,17. Hu et al.18 employed UHR-OCT to produce thickness indices that demonstrated significant diagnostic effectiveness in differentiating between suspected and subclinical KC in comparison to normal eyes. However, there is limited diagnostic power in differentiating between suspected and subclinical KC, and there is insufficient research available to predict the probability of suspected KC advancing to subclinical or even KC. To thoroughly investigate the differences between suspected KC and subclinical KC, as well as to attempt to construct models of the correlation between them for prediction analysis, we extracted all of the detection-related original data from the Pentcam system and the Corvis ST system for this study.

Patients and methods

This study was approved by the Ethics Committee of Qingdao Eye Hospital of Shandong First Medical University. Approval was obtained following the principles outlined in the Helsinki Declaration. Each participant provided informed consent before their data were analyzed.

Inclusion and exclusion criteria for participants

Patients from the Qingdao Eye Hospital of Shandong First Medical University were recruited for this study. The criteria for patients in the suspected KC group included the following: normal slit-lamp examination (stromal thinning, Vogt’s striae, Fleischer’s ring); corrected distance visual acuity of 20/20 or better (Snellen); I-S Value > 1.4D and ≤ 1.9D2,19; presence of yellow or red color-coded number (≥ 1.6 SD) in at least one of the five differential parameters (BADDf, BADDb, BADDp, BADDt, BADDa) in the Belin/Ambrósio Enhanced Ectasia Display; or white or yellow color-coded number in the final parameter “BADD”18; and follow-up for at least one year with no KC development. The criteria for inclusion of patients with subclinical KC were consistent with the abovementioned diagnostic criteria for suspected KC in terms of corrected distance visual acuity, slit-lamp examination, I-S Value, and topographic signs, with KC in fellow eyes3.

The exclusion criteria included individuals who had worn corneal contact lenses for less than 4 weeks (for hard contact lenses) or less than 2 weeks (for soft contact lenses); individuals with a history of eye diseases such as cataracts, retinal detachment, or mild corneal endothelial dystrophy; individuals with moderate to severe dry eyes or high intraocular pressure (IOP); individuals with scar constitution; and individuals who were pregnant or lactating.

Ophthalmology-related examination

The basic data of the patients, including their sex and age, were collected.

Corneal morphological parameters: A total of fifty-five original parameters (using internal CSV files) were collected via the Belin/Ambrósio system and the Topometric/KC system via the Pentacam anterior segment analysis system (Oculus, Germany). These included fifteen Belin/Ambrósio system parameters, such as K1, K2, Astigm, BADDf, BADDb, BADDp, BADDt, BADDa, BADDy, BADD, PachyProgIndexMin, PachyProgIndexMax, PachyProgIndexAvg, ARTMax, DistApexThinLoc; 40 Topometric/KC system parameters: CorneaDia, RSagMin, ISV, IVA, KI, CKI, IHA, IHD, D0 mmPachy, D2 mmPachy, D4 mmPachy, D6 mmPachy, D8 mmPachy, D2 mmStdDev, D4 mmStdDev, D6 mmStdDev, D8 mmStdDev, CVolD3 mm, CVolD5 mm, and CVol. D7 mm, KMaxFront, MaxZonalMean3 mm, KMaxZonalMean4 mm, KMaxZonalMean5 mm, I-S Value, PRC3 mmZone, ARC3 mmZone, ThinnestPachy, ChordMu.polarvector, ABCDA, ABCDB, ABCDC, IrisConvexityMid, IrisConvexityMax, IrisConvexityAxisMax, IrisConvexityMin, IrisConvexityAxisMin, Rf, Rs, CFSagZonalMean4 mm.

The corneal biomechanical parameters included the following: the biomechanical parameters of the cornea were assessed via a Corvis ST Corneal Biomechanics Meter (Oculus, Wegzal, Germany) following a 20 min Pentacam examination for each patient. Thirty-five parameters were collected from the system’s original data (internal CSV file), including A1Time, A1Velocity, A2Time, A2Velocity, HCTime, PeakDist, A1DeformationAmp, A2DeformationAmp, A1DeflectionLength, HCDeformationAmp, HCDeflectionLength, A2DeflectionLength, A1DeflectionAmp, HCDeflectionAmp, A2DeflectionAmp, DeflectionAmpMax_mm, DeflectionAmpMax_ms, A1DeflectionArea, HCDeflectionArea, A2DeflectionArea, A1dArcLength, HCdArcLength, A2dArcLength, dArcLengthMax, MaxInverseRadius, Radius, PachySlope, DefAmpMax, DARatioMax2 mm, IntegratedRadius, DARatioMax1 mm, ARTh, SPA1, CBI.

Statistical methods

All continuous variables were assessed for normality via the Shapiro‒Wilk test. Normally distributed variables were reported as the means ± SDs (x ± s), with group comparisons performed via t-tests. Skewed variables were described by the median and interquartile range (M (P25, P75)), and differences between groups were analyzed by the Wilcoxon rank sum test. Categorical variables were presented as constituent ratios (%), with chi-square tests used for intergroup comparisons.

ROC analysis was performed on various parameters of the Belin/Ambrósio system, Topometric/KC system, and Corvis ST system for the suspected KC group and the subclinical KC group. The best cutoff value was selected as the most approximate index.

After the continuous variables from the three systems were normalized and the categorical variables were encoded into dummy variables, a mathematical prediction model was established via LASSO regression to differentiate between the suspected KC and subclinical KC populations. Through 10-fold cross-validation, the λ with the minimum mean error and λ (1 se) were obtained. A logistic regression-based mathematical prediction model was constructed based on the parameters selected by λ (1 se). The diagnostic performance of the mathematical model was evaluated via the area under the receiver operating characteristic curve (AUC), accuracy, sensitivity, and specificity.

All the statistical analyses were carried out via R studio (version 4.4.0), and the difference was determined as statistically significant when P < 0.05.

Results

Analysis of basic patient data

A total of 144 persons (144 eyes) were included in this study.

Approximately 72 (62.1%) of the 116 individuals (116 eyes) in the suspected KC group were female, and 44 (37.9%) were male, with an average age of 25.00 (20.00, 29.00). The subclinical KC group consisted of 28 people (28 eyes), with an average age of 21.00 (17.75, 24.00) years. Nine (32.1%) of them were female, and 19 (67.9%) were male.

Parameter difference analysis of the two groups in the Belin/Ambrósio system, the Topometric/KC system, and the Corvis ST system

Eleven of the 15 (73.3%) Belin/Ambrósio system parameters (P < 0.05) were significantly different between the suspected KC group and the subclinical KC group, including BADDb, BADDp, BADDt, BADDa, BADDy, BADD, PachyProgIndexMin, PachyProgIndexMax, PachyProgIndexAvg, ARTMax, and DistApexThinLoc. Furthermore, 19 out of 40 (47.5%) Topological/KC system parameters (P < 0.05), including Cornea Diameter, IVA, IHA, IHD, D0mmPachy, D2mmPachy, D4mmPachy, D8mmPachy, D2mmStdDev, D4mmStdDev, D6mmStdDev, D8mmStdDev, CVolD3mm, CVolD5mm, CVolD7mm, PRC3mm.Zone, ThinnestPachy, ABCDB, and ABCDC, were significantly different between the suspected KC group and the subclinical KC group. Finally, 9 out of 34 (26.5%) Corvis ST system parameters (P < 0.05) were significantly different between the suspected KC group and the subclinical KC group, including A1Time, A1DeformationAmp, DARatioMax_2mm, Integrated Radius, DARatioMax_1mm, ARTh, SPA1, and CBI (Table 1).

Table 1 Differenence of all parameters between the subclincial KC group and suspected KC group.

Assessment of the diagnostic power of the two groups under the AUC of the Belin/Ambrósio system, Topometric/KC system, and Corvis ST system

The Belin/Ambrósio system was used to differentiate between the suspected KC group and the subclinical KC group, and 8 out of 15 parameters had an AUC greater than 0.7 (P < 0.05). These parameters included PachyProgIndexMax [0.896 (0.839, 0.954, cutoff = 1.445], ARTMax [0.894 (0.838, 0.951), cutoff = 377.000], BADDa [0.893 (0.836, 0.950), cutoff = 1.010], BADD [0.878 (0.817, 0.939), cutoff = 1.980], BADDp [0.811 (0.730, 0.892), cutoff = 1.935], PachyProgIndexAvg [0.811 (0.730, 0.892), cutoff = 1.185], BADDy [0.758 (0.667, 0.848), cutoff = 0.745], BADDb [0.747 (0.648, 0.847), cutoff = 0.525]. Moreover, we found that 7 of the 40 Topometric/KC system parameters had an AUC greater than 0.7 (P < 0.05), including the D6mmStdDev. Finally, 4 of the 35 Corvis ST system parameters had an AUC greater than 0.7 (P < 0.05), including SPA1 [0.811 (0.718, 0.904), cutoff = 87.485], DARatioMax_2 mm [0.784 (0.691, 0.878), cutoff = 4.710], CBI [0.771 (0.657, 0.885), cutoff = 0.278], DARatioMax_1 mm [0.750 (0.653, 0.847), cutoff = 1.584] (Table 2; Fig. 2).

Table 2 Assessment of the diagnostic ability of the two groups under the AUROC of the three systems.
Fig. 2
Fig. 2
Full size image

ROC curve of Belin/Ambrósio system, Topometric/KC system and Corvis ST system between the two groups.

Screening of sensitive variables via LASSO Regression and Establishment of a logistic regression prediction model

The characteristics of the suspected and subclinical KC populations were screened via LASSO regression. All quantitative parameters were normalized, and categorical variables were set as dummy variables. Variables were screened, and a LASSO regression was created via tenfold cross-validation. Then, λ_min = 0.009 and λ_1se = 0.064 were obtained. We then used λ_1se to construct a LASSO regression and eliminated the following variables: BADDy, BADD, Cornea Diameter, IHD, DARatioMax_2 mm, CBI, and PachyProgIndexMax (Fig. 3; Table 3).

Fig. 3
Fig. 3
Full size image

Coefficient path plot of variables screened by LASSO regression.

Table 3 Variables and coefficients selected by the LASSO regression.

The AUC (95% CI) of the LASSO regression based on the variables listed above was 0.954 (0.921, 0.986); the sensitivity was 0.905; the specificity was 1.000; and the accuracy was 0.908.

Based on the variables selected by LASSO regression, the following logistic regression equation was established:

$$\begin{aligned}logit\left(Risk\:Score\right)&=log(p/(1-p\left)\right)=-62.18+243.19\times\:IHD+3.06\times\:Cornea\:Diameter+2.62\\ & \times\:DARatioMax\_2mm+1.29\times\:BADDy+1.23\times\:BADD+1.03\\ & \times\:PachyProgIndexMax+0.68\times\:CBI\end{aligned}$$

Suspected and subclinical KC were assessed via the risk score derived from the logistic model; the AUC (95% CI) was 0.964 (0.938, 0.991). The sensitivity and specificity were 0.888 and 0.963, respectively, and the accuracy was 0.902 (Fig. 4).

Fig. 4
Fig. 4
Full size image

ROC curve of the logistic regression model.

Discussion

Epidemiological research indicates that the incidence of keratoconus in the global population affects approximately 0.05% of people20. The onset usually occurs at or around the age of 20. Males in Asian populations typically experience disease onset earlier and experience a faster rate of progression than females21, which is in line with the findings from the subclinical KC group in this study.

In this study, we found that, even within the same asymmetrical corneal population, there were morphological and biomechanical differences between the suspected and subclinical KC groups. When comparing the fraction of parameter differences, Pentcam outperformed Corvis ST, particularly for the Belin/Ambrósio system (73.3%). The sensitivity of a single parameter in each system was not very high. PachyProgIndexMax (0.896) and ARTMax (0.894) were the two largest AUCs of the parameters in the Belin/Ambrosio System. These values were higher than those reported in studies conducted by Awad et al.22 and Orkun Muftuoglu. et al.8 compared forme fruste keratoconus eyes to normal eyes (0.64, 0.88; 0.679, 0.772), but the results were lower than those of Vázquez et al. (0.92, 0.93)23 because the majority (85.7%) of the control group they chose had normal map morphology.

Additionally, we discovered that the Belin/Ambrósio deviation correlation parameters (BADDa, BADDb, BADDp, BADDy, and BADD) also had high diagnostic values (AUC = 0.893, 0.747, 0.811, 0.745). It was previously acknowledged that a BADD > 1.6 SD was abnormal23. However, this study discovered that to differentiate between the suspected and subclinical KC groups, BADD > 1.98 SD was considered abnormal. The primary cause of this discrepancy is due to previous studies comparing corneas with normal eyes16,24. The control group selected in this study had asymmetric subclinical KC, and the BADD in some cases was observed to be more than 1.6 SD, which may be the reason for the high BADD cutoff value in this study.

According to our analysis of the Topometric/KC system, IHD, and IVA were the most correlated parameters in terms of marginal corneal dilatation, KC, and irregular astigmatism (AUC = 0.787, 0.762). The greater the degree of vertical asymmetry of the cornea, the greater the likelihood of KC development25,26. Both IHD and IVA significantly increased the progression of subclinical KC (AUC = 0.855, 0.842) after one year of follow-up. et al.27. This finding was consistent with that of the present study, suggesting that IHD and IVA could be used as crucial markers of the development of KC.

The corneal diameter in the subclinical KC group (11.88 ± 0.34) was surprisingly greater than that in the suspected KC group (11.56 ± 0.38). With a lower sensitivity (0.607) but a higher specificity (0.784), the AUC was 0.736 (0.641, 0.832), and the cutoff value was 11.825. Currently, there are significant differences in corneal diameter measurements from country to country. Alfonso JF. et al.28 measured the corneal diameter of myopic individuals via Orbscan TM and reported that it ranged from 10.40 to 13.60 mm, whereas it ranged from 11.50 to 12.30 mm for emmetropia individuals found by Sanchis-Gimeno, J. A et al.29 The corneal diameter of healthy individuals measured with the IOL Master 700 was 11.72 ± 0.45 mm. The corneal diameter of the Chinese myopic population was measured by Xu G. et al.30 via Pentacam, and the result was 11.65 ± 0.38 mm, which was consistent with the results from the suspected KC group (11.56 ± 0.38 mm) in our study. The corneal diameter was found to be greater in both KC and contralateral eyes (12.14 mm and 12.17 mm), according to research by François-Xavier Crahay et al.31. The corneal diameter we measured in the subclinical KC group was not as large as that measured by François-Xavier Crahay because the corneal diameter and shape of the white population differ from those of the Chinese population32,33. However, the KC population we studied had a larger corneal diameter. Thus, corneal diameter may be a crucial parameter for differentiating corneas in KC patients from those in the normal group.

The use of biomechanical examination is still vital in differentiating between KC and healthy corneas16,34. In this study, we found that 26.5% of the Corvis ST system parameters differed between the two groups. The highest AUC values were found for SPA1 (0.811), DARatioMax_2mm (0.784), CBI (0.771), and DARatioMax_1mm (0.750). No parameter had a particularly high diagnostic efficacy, but this is subject to future research.

Racial differences in ocular parameters are critical for disease diagnosis, extending beyond corneal diameter to biomechanical metrics. Research conducted by Liu, Yan et al.35 and Guo, Li-Li et al.36 documented elevated false-positive rates of the CBI within Chinese populations. Vinciguerra, Riccardo et al.37 highlighted disparities in SPA1, SSI, and ARTH between Chinese and Caucasian populations, subsequently introducing the cCBI, which outperformed CBI in diagnosing keratoconus in Chinese individuals. Additionally, Guo, Yihan et al.38 revealed variations in the age-biomechanics relationship between these populations. This study focuses on Chinese samples, limiting its applicability to other ethnicities. Future multicenter, multiethnic studies are needed to confirm these findings.

No parameter in our work has an AUC of more than 0.9 because KC, especially in its early stages, cannot be diagnosed with a single parameter; rather, a comprehensive analysis necessitates the combination of multiple characteristics. Therefore, we developed a LASSO regression prediction model, which was first suggested in 1996 by Robert Tibshirani39. This method of compressed estimation is used here. It produces a more refined model by building a penalty function that compresses some coefficients and sets others to zero. Complex collinear data are specifically designed to be handled by this biased estimation. There were 92 parameters in this study; however, there may have been connections between the parameters. We chose seven parameters via LASSO regression. AUC = 0.964 indicates that the logistic regression model based on the seven parameters has increased diagnostic power, effectively resolving the issue that prior studies were unable to differentiate between suspected and subclinical KC18.

Currently, a large body of research using artificial intelligence to predict the progression of keratoconus has been conducted14,15,17,40,41, with good diagnostic power. For the first time, we used the internal parameters of the examination system to compare the differences between the two types of asymmetrical corneas in our study, and the result was a more precise prediction model. The results of our analysis of the samples we used in this model showed that IHD had the largest coefficient, indicating the strongest correlation with the development of the KC. We then studied the corneal diameter, DARatioMax_2mm, BADDy, BADD, PachyProgIndexMax, and CBI. The logistic regression model’s AUC was 0.964, which was greater than that of studies of sequences 1 (0.849), 2 (0.950), and 4 (0.945). This study’s sensitivity (0.963) was greater than that of sequences 1–4 (0.754, 0.903, 0.936, 0.844), but its specificity (0.888) was lower than that of sequences 2–4 (0.943, 0.972, 0.901) (Table 4).

Table 4 Classification, methods and AUC of previous relevant studies and this study.

The sample sizes for all of the abnormal corneas were greater than those used in this investigation above researches. Importantly, this study’s sample size was limited, which might have had an impact on the model’s capacity for accurate prediction. Furthermore, biased estimation may arise if the sample data do not accurately reflect the characteristics of the real subclinical and suspected populations. Importantly, the model might not precisely encapsulate the attributes of the complete population. In future investigations, we will gather additional clinical practice samples to improve our model and increase the authenticity and accuracy of model prediction.

In conclusion, the probability of developing keratoconus in suspected cases may be influenced by a combination of morphological and biomechanical factors in Chinese populations. Evaluating parameters such as increased corneal diameter, Belin-related indices, and biomechanical metrics could be instrumental in determining the likelihood of keratoconus, especially in the context of inferior and superior corneal asymmetry.Integrating these factors into the assessment process could improve safety and enhance decision-making in refractive surgery.