Abstract
Infections associated with cardiac implantable electronic devices (CIEDs) pose significant clinical challenges due to their life-threatening nature and complex management. Recent advancements in prevention, diagnosis, and treatment have been driven by landmark trials such as PADIT and WRAP-IT, alongside evolving diagnostic tools like [18 F]FDG PET/CT and updated diagnostic criteria (e.g., 2023 Duke-ISCVID). A combination of well-established approaches—such as leadless pacemakers, subcutaneous ICDs, and an antibiotic-eluting envelope—and evolving adjunctive strategies, including a taurolidine-containing antimicrobial agent, incision drapes, fascial plane blocks or double gloving, have broadened the spectrum of options to reduce and potentially mitigate the risk of CIED infection. While devices like leadless pacemakers, subcutaneous ICDs, and antibiotic envelopes are supported by randomized controlled trials, adjunctive measures such as taurolidine rely predominantly on observational data. To address this variation in the strength of available evidence, a modified Delphi consensus process was conducted, bringing together cardiologists and infectious disease experts to define experience-based best practices in areas where current guidelines offer limited or no specific recommendations. The consensus achieved strong agreement (≥ 80%) on key strategies, including mandatory double-gloving to reduce contamination, the use of validated risk models (e.g., PADIT, BLISTER) for tailored infection prevention employing an antibiotic eluting envelope, multidisciplinary decision-making for non-extractable infections and the use of taurolidine solutions as adjuncts to prevent CIED infection. Moderate consensus (60–79%) supported non-delayed reimplantation after lead-related endocarditis and single-session reimplantation in pacing-dependent patients under specific circumstances. However, no consensus was reached on mandating a minimum annual procedural volume (≥ 500 cases) for CIED centers, reflecting concerns about access and operator expertise. The findings emphasize the importance of personalized risk stratification, procedural innovations, and multidisciplinary collaboration in optimizing CIED infection outcomes. Areas requiring further research include the efficacy of iodophor-impregnated drapes, fascial plane blocks, and taurolidine solutions. This consensus provides a pragmatic framework for clinicians, highlighting evidence-informed strategies to mitigate infection risks and improve care for patients undergoing CIED procedures.
Similar content being viewed by others
Introduction
Infections associated with cardiac implantable electronic devices (CIEDs) represent a significant clinical challenge due to their life-threatening nature and the complexities involved in both diagnosis and management. Among these, lead-related infective endocarditis (L-IE) is a particularly problematic complication, often complicating clinical decision-making and patient outcomes. Since the initial scientific statements issued by key professional bodies such as ECAS, EHRA, AHA, HRS, and CTS, substantial advancements have been made in the prevention, diagnosis, and treatment of CIED infections.1,2,3,4,5,6.
Notably, landmark randomized controlled trials like PADIT7 and WRAP-IT8 have provided pivotal evidence supporting infection prevention strategies, including the use of antibiotic-impregnated envelopes during device implantation. Diagnostic modalities have also evolved, with critical reassessment of transesophageal echocardiography (TEE) revealing limitations in differentiating infectious from noninfectious lead masses, and the growing utilization of fluorine-18-fludeoxyglucose positron emission tomography/computed tomography ([18 F]FDG PET/CT) enhancing diagnostic accuracy.2 The recent update to the 2023 Duke-International Society for Cardiovascular Infectious Diseases criteria further recognizes CIED presence as a specific minor criterion for L-IE diagnosis, reflecting its established role as a predisposing factor.4.
In parallel, novel device technologies and adjunctive therapies, including Taurolidine-containing antimicrobial agents9,10 leadless pacemakers (LPMs),5 extravascular implantable cardioverter defibrillators (EV-ICDs)11 and subcutaneous implantable cardioverter defibrillators (S-ICDs)12, have expanded therapeutic options. Infection risk scores have gained traction as tools to stratify patients and tailor preventive measures, guiding both device selection and the application of infection mitigation strategies such as antibiotic eluting envelopes13. Additionally, emerging techniques like percutaneous mechanical aspiration systems offer minimally invasive approaches to managing complicated right-sided vegetations in select cases14.
Building on these advances, this Delphi consensus statement15 was developed by an expert panel committed to improving outcomes for patients undergoing CIED procedures through the promotion of evidence-informed best practices aimed at minimizing infection risk. While this statement does not constitute a comprehensive review of all facets of CIED infection, including detailed clinical features, epidemiology, and pathogenesis, nor does it serve as formal clinical guidelines, it provides pragmatic strategies and algorithms designed to support clinicians worldwide in optimizing care for this vulnerable patient population.
Methods
To develop a consensus statement on the management and prevention of CIED infections, a modified Delphi method was conducted in accordance with the “Reporting guidelines for Delphi techniques in health sciences”.16. The expert panel comprised faculty from the 360° CIED Infection Congress, including cardiologists, cardiothoracic surgeons and infectious disease specialists practicing in Italy, Germany, Austria, and the United Kingdom, representing tertiary referral centers with extensive clinical experience and published expertise in device-related infections, who gave consent deliberately to participate in the survey before the survey and during the congress itself.
Invited experts participated in iterative survey rounds conducted anonymously via SurveyMonkey® (SVMK, San Mateo, CA, USA), aiming to achieve consensus on the most critical topics in CIED infection prevention and management.
Several measures were implemented to reduce bias. Participation was voluntary but restricted to clinicians with established expertise in CIED infections. All voting rounds were conducted anonymously, preventing influence from dominant personalities or hierarchical relationships within the panel. Aggregated results were shared between rounds to allow participants to reconsider their responses while maintaining anonymity.
The modified Delphi process consisted of three sequential stages. First, during the 360° CIED Infection Congress, panel members participated in an exploratory round in which key clinical areas lacking clear guideline recommendations were identified and discussed. Based on this discussion, the coordinating group formulated ten preliminary statements reflecting the topics considered most relevant to CIED infection prevention and management.
In the second stage (first formal voting round), panelists rated each statement using a 5-point Likert scale (1 = strongly agree, 5 = strongly disagree). Following this round, aggregated voting results were circulated to participants. The strength of agreement was classified as strong agreement (≥ 80%), moderate agreement (> 55%) or no agreement (< 50%) as extrapolated from the RAND methodology. The results were presented via electronic mail to all participants.
In the third stage (final consensus round), panelists were invited to reconsider the statements in light of the group responses from the first formal voting round. Statements were not substantially modified between rounds, but clarifications were provided where wording had led to uncertainty. The results from this final round were used for the consensus analysis presented in this manuscript. In addition to reporting overall agreement rates (scores 1–2), the distribution of responses across the full 5-point Likert scale was recorded to enhance transparency of voting patterns (Table S1).
Online consensus rounds were conducted anonymously. The results were exported from SurveyMonkey® and analyzed using Microsoft Excel (Microsoft Corporation, Redmond, USA).
In accordance with EU GDPR Recital 26 and the guidance of the Italian “Garante per la Protezione dei Dati Personali”, the data collected qualify as fully anonymized and therefore fall outside the scope of data protection legislation granting a waiver of formal ethical approval.
Results
Of the 21 members of the 360° CIED Infection Faculty, 20 participated in the final round of the modified Delphi consensus process. Consensus was achieved in nine out of ten questions. The questions and corresponding results are summarized in Table 1. The full distribution of Likert responses for each statement is presented in Table S1 in the Supplementary to provide greater transparency regarding the degree of agreement and dissent among panel members.
Discussion
Double-gloving protocol (Q1)
A strong consensus (81.25% agreement) supports mandatory double-gloving with glove changes post-draping to minimize contamination risk. Double-gloving—wearing two pairs of sterile surgical gloves—has been extensively studied for its role in reducing glove perforation rates and enhancing infection control during surgical procedures. A comprehensive Cochrane review analyzed 14 trials and found that double-gloving significantly reduces perforations to the innermost gloves, with an odds ratio of 4.10 (95% CI 3.30 to 5.09) compared to single gloves.17 This reduction in glove breaches is crucial for minimizing the risk of pathogen transmission between the surgical team and the patient. Further supporting these findings, a systematic review encompassing 18 randomized controlled trials reported that double-gloving decreases blood contamination risk by 65% and inner glove perforation by 71% compared to single gloves.18 Additionally, the use of indicator gloves—colored latex gloves worn underneath standard gloves—has been shown to significantly improve the detection of perforations during surgery, enhancing the ability to maintain sterility.19 Despite these advantages, the adoption of double-gloving is not universal in all surgical disciplines, including CIED procedures. Concerns regarding manual dexterity and increased resource utilization are valid; however, studies suggest that double-gloving does not significantly impair surgical performance. Implementing double-gloving protocols may require adjustments in training and workflow to ensure that the benefits outweigh the costs.
Use of risk models (Q2)
With 88.23% agreement, there is strong endorsement for employing validated infection risk prediction models such as PADIT,7 BLISTER,13 or SHARIFF20 to guide antibiotic envelope use and tailor prophylaxis. These models reflect an evolving shift toward personalized medicine in CIED infection prevention, enabling stratification of patients by risk and more judicious antibiotic use, which is critical to mitigate antimicrobial resistance. Current studies support these tools’ predictive accuracy, yet their clinical utility depends on ease of integration into routine practice and ongoing validation across diverse patient cohorts.21.
Alternative treatment pathways in non-extractable infections (Q3)
The consensus (88.23% agreement) favors heart team–led multidisciplinary decision-making incorporating patient and family preferences for managing complex CIED infections where device extraction is not feasible. Alternative approaches like targeted ultra-high in-situ antibiotic administration22 or capsulectomy, thorough debridement and taurolidine-containing solution antisepsis with subpectoral repositioning23,24,25,26,27 align with emerging clinical practice emphasizing individualized care plans, especially in frail or anatomically challenging patients. This paradigm prioritizes balancing infection control with quality of life and procedural risk, as supported by contemporary case series and observational data. The panel noted that this statement encompassed several complex clinical scenarios (frailty, non-extractable systems, and alternative infection management strategies). While agreement remained strong, the breadth of the statement may limit interpretability and highlights the need for future work to address these scenarios individually.
Reimplantation after lead-related endocarditis (Q4)
A reasonable consensus (70.59% agreement) supports device reimplantation as early as one week following blood culture clearance in lead-related endocarditis cases. This aligns with current infectious disease guidelines that recommend ensuring infection resolution prior to reimplantation to reduce relapse risk, but lack clear guidance on timing.4,5 The minority expressing neutral or dissenting views reflect variability in institutional protocols or risk tolerance, underscoring the need for additional prospective studies to optimize courageous timing supported here.
Single-session reimplantation after local infection (Q5)
Support for contralateral single-session reimplantation in pacing-dependent patients (70.59% agreement) reflects clinical pragmatism when strict sterile conditions and complete microbiological workup are ensured. However, a notable neutral rate (17.65%) suggests ongoing debate regarding the balance between infection control and urgent pacing needs. Given the limited existing evidence, it’s crucial to prioritize individualized assessments including proactive prevention strategies, and adherence to rigorous infection screening.
Minimum center volume requirement (Q6)
Disagreement predominates (47.05% vs. 41.18% agreement) regarding a minimum annual procedural volume (≥ 500 CIED cases) for centers performing procedures such as CIED implants, revisions and extractions. The lack of consensus reflects complex factors: operator skill and team experience may outweigh absolute procedural numbers, and arbitrary cutoffs risk limiting access to care. Literature indicates volume–outcome relationships but fails to define precise thresholds, highlighting the need for nuanced quality metrics beyond volume alone.5.
Routine use of fascial plane blocks (Q7)
Moderate support (58.82% agreement, 29.41% neutral) exists for fascial nerve blocks as adjuncts to improve pain control and potentially reduce infection risk through decreased wound manipulation.28 While promising data from analogous surgical disciplines exist, specific evidence in CIED populations remains limited. Training requirements and procedural logistics further temper enthusiasm, pointing to the need for focused clinical trials to define safety, efficacy, and standardized protocols.
Iodophor-impregnated drapes (Q8)
Moderate support (58.82% agreement, 35.29% neutral) was noted for iodophor-impregnated drapes. These antimicrobial drapes have demonstrated efficacy in reducing surgical site infections across multiple specialties, attributed to sustained local antiseptic action. However, neutrality among panelists likely reflects concerns about increased costs, resource implications, and inconsistent evidence specific to CIED procedures when employed adjunctively to standard sterile draping. While cost-effectiveness analyses are sparse, available data suggest potential benefit, warranting further large-scale comparative trials to clarify their role and facilitate cost-benefit assessments in this specific setting.29.
Use of taurolidine solutions (Q9)
Strong consensus (88.24%) endorses taurolidine-containing antimicrobial agents as valuable adjuncts in infection prevention. Taurolidine is a synthetic taurine derivative with broad antimicrobial activity against gram-positive, gram-negative, and fungal organisms, as well as anti-biofilm properties. It has been widely used in Europe in the prevention of catheter-related infections and has more recently been investigated as an adjunctive antiseptic strategy during CIED procedures9,10 Despite promising results, the panel appropriately calls for more data to confirm definitive clinical efficacy before recommendation in all CIED procedures.
Patient fixation during implantation (Q10)
Although seemingly procedural in nature, patient stabilization during device implantation was included because unintended patient movement may compromise sterile conditions and procedural safety, particularly in procedures performed under conscious sedation. Reasonable support (64.71% agreement) exists for routine use of patient strap fixation to reduce patient movement during CIED procedures ideally minimizing lead dislodgement and contamination risk. While patient fixation during CIED procedures is widely regarded as a best practice, some variability in neutral responses suggests differences in procedural habits or patient-specific considerations. Current evidence from observational studies supports strap fixation as a simple, effective measure to enhance procedural outcomes and should be further evaluated in controlled studies.
Limitations
Several limitations inherent to the Delphi methodology should be acknowledged. First, although the panel included cardiologists, cardiac surgeons, and infectious disease specialists with recognized expertise in CIED infections, the expert group was relatively small and consisted primarily of European clinicians, which may limit the generalizability of the findings to other healthcare systems and practice environments.
Second, Delphi consensus reflects expert opinion rather than direct clinical evidence and may therefore be influenced by prevailing clinical experience and local practice patterns. While anonymity of voting was maintained to reduce peer influence and dominance bias, residual bias cannot be fully excluded.
Third, the statements were generated by the expert panel itself, which may introduce topic selection bias toward areas perceived as clinically relevant by the panel rather than representing all unresolved issues in CIED infection management.
Fourth, the consensus process does not replace guideline development based on systematic evidence review, but instead provides experience-based guidance in areas where robust clinical trial data remain limited.
Finally: Practice patterns, resource availability, and regulatory environments vary substantially between regions; therefore, some recommendations discussed here may not be directly transferable to all healthcare systems.
Conclusion
This consensus statement reflects the collective opinion of a multidisciplinary panel of experts regarding practical approaches to CIED infection prevention and management in areas where guideline recommendations remain limited. Through a systematic exploration of critical topics, including procedural best practices, patient-specific risk stratification, emerging prophylactic adjuncts, and institutional quality benchmarks, this consensus seeks to advance clinical decision-making with strategies that are both evidence-informed and pragmatically applicable. Importantly, areas of only moderate agreement, such as the use of iodophor-impregnated drapes and fascial plane blocks, reflect ongoing uncertainties and underscore the need for targeted research to refine and validate these practices. This evolving body of knowledge highlights the dynamic nature of CIED infection prevention and the necessity of continual reassessment as new data emerge.
Data availability
All relevant data generated or analyzed during this study are included within this manuscript. No additional datasets were generated or used.
References
Blomstrom-Lundqvist, C. & Ostrowska, B. Prevention of cardiac implantable electronic device infections: guidelines and conventional prophylaxis. Europace 23, iv11–19. https://doi.org/10.1093/europace/euab071 (2021).
Blomstrom-Lundqvist, C. et al. European Heart Rhythm Association (EHRA) international consensus document on how to prevent, diagnose, and treat cardiac implantable electronic device infections-endorsed by the Heart Rhythm Society (HRS), the Asia Pacific Heart Rhythm Society (APHRS), the Latin American Heart Rhythm Society (LAHRS), International Society for Cardiovascular Infectious Diseases (ISCVID), and the European Society of Clinical Microbiology and Infectious Diseases (ESCMID) in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS). Eur. Heart J. https://doi.org/10.1093/eurheartj/ehaa010 (2020).
Traykov, V. et al. Clinical practice and implementation of guidelines for the prevention, diagnosis and management of cardiac implantable electronic device infections: results of a worldwide survey under the auspices of the European Heart Rhythm Association. Europace https://doi.org/10.1093/europace/euz137 (2019).
Fowler, V. G. et al. The 2023 Duke-international society for cardiovascular infectious diseases criteria for infective endocarditis: updating the modified duke criteria. Clin. Infect. Dis. 77, 518–526. https://doi.org/10.1093/cid/ciad271 (2023).
Baddour, L. M. et al. Update on Cardiovascular Implantable Electronic Device Infections and Their Prevention, Diagnosis, and Management: A Scientific Statement From the American Heart Association: Endorsed by the International Society for Cardiovascular Infectious Diseases. Circulation 149, e201–e216. https://doi.org/10.1161/CIR.0000000000001187 (2024).
Kusumoto, F. M. et al. 2017 HRS expert consensus statement on cardiovascular implantable electronic device lead management and extraction. Heart Rhythm. 14, e503–e551. https://doi.org/10.1016/j.hrthm.2017.09.001 (2017).
Krahn, A. D. et al. Prevention of Arrhythmia Device Infection Trial: The PADIT Trial. J. Am. Coll. Cardiol. 72, 3098–3109. https://doi.org/10.1016/j.jacc.2018.09.068 (2018).
Tarakji, K. G. et al. Antibacterial envelope to prevent cardiac implantable device infection. N Engl. J. Med. https://doi.org/10.1056/NEJMoa1901111 (2019).
Vonthein, R. et al. Taurolidine-containing solution for reducing cardiac implantable electronic device infection-early report from the European TauroPace™ registry. J. Cardiothorac. Surg. 19, 592. https://doi.org/10.1186/s13019-024-03059-1 (2024).
Borov, S. et al. Use of a taurolidine containing antimicrobial wash to reduce cardiac implantable electronic device infection. EP Europace. https://doi.org/10.1093/europace/euad306 (2023).
Friedman, P. et al. Efficacy and safety of an extravascular implantable cardioverter-defibrillator. N Engl. J. Med. 387, 1292–1302. https://doi.org/10.1056/NEJMoa2206485 (2022).
Knops, R. E. et al. Subcutaneous or transvenous defibrillator therapy. N Engl. J. Med. 383, 526–536. https://doi.org/10.1056/NEJMoa1915932 (2020).
Maclean, E. et al. The BLISTER Score: A Novel, Externally Validated Tool for Predicting Cardiac Implantable Electronic Device Infections, and Its Cost-Utility Implications for Antimicrobial Envelope Use. Circ. Arrhythm. Electrophysiol. 17, e012446. https://doi.org/10.1161/circep.123.012446 (2024).
Moriarty, J. M. et al. Endovascular Removal of Thrombus and Right Heart Masses Using the AngioVac System: Results of 234 Patients from the Prospective, Multicenter Registry of AngioVac Procedures in Detail (RAPID). J. Vasc Interv Radiol. 32, 549–557e543. https://doi.org/10.1016/j.jvir.2020.09.012 (2021).
Boulkedid, R., Abdoul, H., Loustau, M., Sibony, O. & Alberti, C. Using and reporting the Delphi method for selecting healthcare quality indicators: a systematic review. PLoS One. 6, e20476. https://doi.org/10.1371/journal.pone.0020476 (2011).
Spranger, J., Homberg, A., Sonnberger, M. & Niederberger, M. Reporting guidelines for Delphi techniques in health sciences: A methodological review. Z. für Evidenz Fortbild. und Qualität im Gesundheitswesen. 172, 1–11. https://doi.org/10.1016/j.zefq.2022.04.025 (2022).
Tanner, J. & Parkinson, H. Double gloving to reduce surgical cross-infection. Cochrane Database Syst. Rev. 2006, Cd003087. https://doi.org/10.1002/14651858.CD003087.pub2 (2006).
Zhang, Z., Gao, X., Ruan, X. & Zheng, B. Effectiveness of double-gloving method on prevention of surgical glove perforations and blood contamination: A systematic review and meta-analysis. J. Adv. Nurs. 77, 3630–3643. https://doi.org/10.1111/jan.14824 (2021).
Meakin, L. B., Gilman, O. P., Parsons, K. J., Burton, N. J. & Langley-Hobbs, S. J. colored indicator undergloves increase the detection of glove perforations by surgeons during small animal orthopedic surgery: a randomized controlled trial. Vet. Surg. 45, 709–714. https://doi.org/10.1111/vsu.12519 (2016).
Shariff, N. et al. Health and economic outcomes associated with use of an antimicrobial envelope as a standard of care for cardiac implantable electronic device implantation. J. Cardiovasc. Electrophysiol. 26, 783–789. https://doi.org/10.1111/jce.12684 (2015).
Ziacchi, M. et al. REducing INFectiOns thRough Cardiac device Envelope: insight from real world data. REINFORCE Project Europace. https://doi.org/10.1093/europace/euad224 (2023).
Topaz, M. et al. Regional antibiotic delivery for implanted cardiovascular electronic device infections. J. Am. Coll. Cardiol. 81, 119–133. https://doi.org/10.1016/j.jacc.2022.10.022 (2023).
Giudice, M. et al. Use of taurolidine in a patient with a cardiac implantable electronic device protrusion. JACC: Case Rep. 14, 101835. https://doi.org/10.1016/j.jaccas.2023.101835 (2023).
Weichsel, J. et al. Eradication of ventricular assist device driveline infection in paediatric patients with taurolidine. J. Cardiovasc. Dev. Disease. 9, 18 (2022).
Borov, S., Baldauf, B., Lau, E. W. & Bonnemeier, H. Salvage of infected cardiac implantable electronic device with taurolidine—a case report. Cardiothorac. Surgeon. 30, 7. https://doi.org/10.1186/s43057-022-00068-5 (2022).
Giaccardi, M., Baldauf, B., Lau, E. W., Borov, S. & Bonnemeier, H. Salvage of cardiac implantable electronic device pocket infection with skin erosion in frail 92-year-old. J. Cardiovasc. Dev. Disease. 9, 81 (2022).
Casorelli, E. et al. [Pacemaker infection in fragile patients]. Herzschrittmacherther Elektrophysiol. https://doi.org/10.1007/s00399-023-00940-9 (2023).
Capuano, P. et al. Fascial plane blocks for cardiothoracic surgery: a narrative review. J. Anesth. Analgesia Crit. Care. 4, 20. https://doi.org/10.1186/s44158-024-00155-5 (2024).
Golian, M. et al. Adherent skin barrier drape use is associated with a reduced risk of cardiac implantable device infection: Results from a prospective study of 14,225 procedures. Heart Rhythm O2. 5, 103–112. https://doi.org/10.1016/j.hroo.2023.12.002 (2024).
Funding
Open Access funding enabled and organized by Projekt DEAL. Open access publishing for this article was supported through a transformative agreement between the DEAL Consortium and Springer Nature. Other, than stated above this research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.
Author information
Authors and Affiliations
Contributions
Conception and design: HBO, BB, MG, RC. Data collection: BB, HBO. Data analysis and interpretation: BB, KB, MG, RC and HBO. Writing of the manuscript draft: BB, KB, RC, HBO, MG. Manuscript review: all authors. Final approval of the manuscript: all authors.
Corresponding author
Ethics declarations
Competing interests
BB is a medical consultant for/part of the advisory board of: Abbott, Bioline Supply, Biotronik, Cablon NL, CRM Microport, Crosstec GmbH, Drugsales Limited, DSE Türkiye, Implicity, Kappamed, Kimal PLC, M3 Medical/Ecclipse Medical, Medival SRL, Medtronic, Philips/Spectranetics, Sintec SRL, Tauro-Implant GmbH, Tauropharm GmbH, Transcutan, Transaxis. The other authors have no conflicts of interest to declareFunding: Open access publishing for this article was supported through a transformative agreement between the DEAL Consortium and Springer Nature.
Ethical approval
All methods were carried out in accordance with relevant guidelines and regulations, including the ethical principles set forth in the latest version of the Declaration of Helsinki. This Delphi consensus study was conducted exclusively among faculty members attending the 360 CIED Infection Congress on May 9th, 2025, in Florence, Italy.The survey was conducted fully anonymously; no personal identifiers, digital signatures, or traceable metadata were collected or stored. In accordance with EU GDPR Recital 26 and the guidance of the “Garante per la Protezione dei Dati Personali”, anonymous data that do not allow for the identification of individuals are not subject to data protection obligations. As such, the survey data qualify as anonymous, and a waiver of formal data protection review is granted. Although all participating faculty members are co-authors of this manuscript, the survey data were collected and analyzed in such a way that individual responses cannot be linked to any participant, thereby maintaining full anonymity throughout the consensus process. All experimental protocols were reviewed for compliance with institutional standards. Informed consent was obtained from all participating faculty members prior to the start of the survey, during the congress itself.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Below is the link to the electronic supplementary material.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Baldauf, B., Bode, K., Biffi, M. et al. Delphi consensus recommendations for preventing and treating cardiac implantable electronic device infections beyond current guidelines. Sci Rep 16, 13135 (2026). https://doi.org/10.1038/s41598-026-49515-x
Received:
Accepted:
Published:
Version of record:
DOI: https://doi.org/10.1038/s41598-026-49515-x