Table 1 Precision study of the Hilab Lens device

From: Clinical validation of the Hilab lens for AI supported point of care CBC testing in pediatric oncology

Analyte

Reference value

Evaluated Parameters

Summary

Expected results*

Average

SD

Assay CV

Random Error

Bias

Assay CV

Bias (Average)

Random Error (Max)

Total Error

CV (95% CI)

Bias (95% CI)

Total Error (95% CI)

Intra

Inter

Intra (Max)

Inter (Max)

RBC (x 106 /mm³)

4.2

4.27

0.12

2.73%

0.96%

4.21%

1.57%

2.78%

1.65%

2.62%

4.82%

8.35%

3.6%

6%

8.5%

2.6

2.65

0.07

2.39%

1.24%

3.99%

2.56%

3.6

3.70

0.11

2.78%

1.65%

4.82%

3.73%

HB (g/dL)

7.8

8.1

0.26

3.10%

0.34%

4.46%

4.36%

3.10%

1.20%

2.30%

4.46%

7.60%

2.7%

4.3%

8.8%

17.4

17.05

0.37

2.08%

0.81%

3.26%

−1.98%

12.2

12.7

0.36

2.71%

1.20%

4.35%

4.23%

PLA (/mm³)

29,800

27,831

1,792

6.48%

1.98%

9.82%

−6.61%

6.48%

5.54%

−1.60%

11.63%

12.22%

7.4%

10.9%

23.1%

211,000

211,153

13,395

6.06%

1.24%

8.90%

0.07%

61,000

62,065

3,225

5.04%

5.54%

11.63%

1.75%

WBC (/mm³)

4,100

4,181

446

10.56%

3.30%

16.04%

1.98%

10.56%

3.30%

3.54%

16.04%

22.59%

11.9%

10.5%

30.2%

6,240

6,587

498

7.37%

0.71%

10.60%

5.56%

16,800

17,319

1145

6.38%

2.19%

9.80%

3.09%

  1. *Expected results based on established parameters by EFLM Biological Variation (higher CI limit).
  2. The table demonstrates values of average, standard deviation (SD), coefficient of variability (CV; Inter and Intra-Assay), Bias, Random Error, and Total Error of each analyte.