Introduction

Peripheral nerve injuries are a prevalent clinical issue, affecting an estimated 13 to 23 individuals per 100,000 annually1. Repairing of severe traumatic nerve injuries continues to be a formidable challenge and the outcomes remain unsatisfactory2. Inadequate treatment can result in permanent motor, sensory, and autonomic impairments. Following nerve transection injuries (e.g., severe injuries or amputation), abnormal nerve tissue regrowth in a disorganized manner often occurs, resulting in traumatic neuroma formation and subsequent chronic pain (as many as 36.5% of patients reported chronic neuropathic pain following traumatic nerve injury in the upper extremity)3,4,5. In cases of amputation, neuroma formation can also greatly complicate prosthetic integration4. This can impact procedures such as targeted muscle reinnervation (TMR), which aims to enhance myoelectric prostheses by transferring transected residual nerves to alternative distant muscles, and disrupted tissue regrowth by neuroma could cause significant difficulties wearing a prosthesis6,7,8,9.

Since early repair and diagnosis is crucial for optimal therapeutic results5, monitoring the initial stages of healing after severe nerve injuries is essential, which can offer vital physiological and pathological insights for timely interventions and enhanced function restoration10,11,12,13. Multifunctional peripheral nerve interfaces present a promising approach for remote and continuous monitoring of nerve injuries, which however remains largely unexplored. By integrating both restorative and monitoring functions, the nerve interface has the potential to promote nerve regeneration and facilitate early detection of recovery challenges during the initial phases of nerve repair. This enables prompt interventions and enhances treatment optimization, ultimately maximizing rehabilitation outcomes and prosthetic integration.

One major challenge in developing neural interfaces for probing nerve injuries is that conventional peripheral nerve interfaces (e.g., nerve cuffs) are based on non-degradable materials14,15, and often necessitate additional surgeries for subsequent device removal to avoid permanent materials retention, imposing increased infection risks and associated hospital costs. In addition, fibrotic encapsulation typically occurs around the implant especially at the injured site due to foreign body reactions16,17, further complicating the extraction procedure and risking secondary damage to the nerve tissues. Recent developments in bioresorbable materials and devices indicate great potential18,19. Building fully biodegradable peripheral nerve interfaces can offer a promising approach to mitigate the issue by obviating the need for surgical retraction. However, existing biodegradable interfaces focus on stimulation for nerve regeneration or pain management, which lacks monitoring capabilities6,20. The development of bioresorbable multifunctional nerve interfaces capable of both neural recording and supporting nerve regrowth remains unexplored. Such a biodegradable electronic platform can enable temporary monitoring and treatment, where early-stage tracking and intervention are crucial to optimize clinical outcomes but become unnecessary once regeneration or reinnervation is complete21.

In this paper, we present a biodegradable and restorative neural interface (Bio-Restor) designed for concurrent monitoring and facilitating the healing of long-gap nerve injuries at a finite time frame matched to the critical initial phase of nerve recovery (Fig. 1). The device consists of an electrode array for nerve electrical recording and a self-powered galvanic cell that delivers electrical cues to support tissue repair. The interface is built upon a flexible and biodegradable polymer substrate, complemented by a self-morphing shape memory polymer film, which aids the surgical placement of the device onto the injured nerve. The biodegradable neural interface is integrated with a wireless bio-potential recorder for signal amplification, digitization, and remote transmission. Employing machine learning techniques, we realize the evaluation of nerve recovery status and the early identification of neuroma formation and accomplish greatly enhanced therapeutic outcomes (Fig. 1). A key attribute of this electronic platform is its construction from fully biodegradable materials, which degrades into biologically benign byproducts after the monitoring window, thus eliminating the need for surgical retraction and reducing the risk of tissue damage and infection. The proposed material strategy and device scheme open up new avenues for the interrogation and intervention of nerve injuries to accomplish optimal therapeutic results.

Fig. 1: Biodegradable and restorative neural interface (Bio-Restor) for wireless interrogation of neuropathic injuries.
figure 1

Schematic illustration of the device, which is composed of a biodegradable substrate PLLA-PTMC (30 μm, 10 × 10 mm), a Mg-FeMn galvanic cell (Mg, 3.5 μm, 1.5 × 5 mm; FeMn, 1.5 μm, 1.5 × 5 mm), Mo/Si recording electrode array (Mo, 300 nm, 0.3 × 5 mm; Si, 2 μm, 0.5 × 5 mm), and a biodegradable shape memory polymer layer (400 μm, 10 × 5 mm). Evaluation of nerve recovery status and early detection of neuroma have been demonstrated based on the biodegradable, restorative, and self-morphing neural interface.

Results

Materials strategies and device fabrication

The illustration of the device structure of Bio-Restor is depicted in Fig. 1, with the corresponding fabrication process given in Supplementary Fig. 1. Specifically, the device includes a 7-channel electrode array for signal recording and a galvanic cell to facilitate functional recovery (Fig. 1). The bilayer dissolvable recording electrodes are composed of molybdenum (Mo) thin films (300 nm) and highly doped N-type monocrystalline silicon (Si) membranes (2 µm thick, conductivity 200–1000 S/cm), to ensure desirable conductivity and sufficient operational lifetime. The underlying rationale is based on the slow dissolution rates of highly doped Si, which can serve as effective and conductive water barriers for Mo electrodes22,23. Mo electrodes are deposited through sputtering by a shadow mask on a biodegradable and flexible substrate which is a copolymer of poly(l-lactic acid) and poly(trimethylene carbonate) (PLLA-PTMC, 60:40 ratio, ~30 μm thick), followed by the transfer-printing of Si membranes to achieve the bilayer electrode array (Supplementary Fig. 1). A biodegradable galvanic cell comprised of thin-film magnesium (Mg, 3.5 µm) and iron-manganese alloy (FeMn, 1.5 µm) electrodes are deposited on the PLLA-PTMC substrate next to the Mo/Si electrode array (Fig. 1 and Supplementary Fig. 1), providing electrical cues to promote nerve regrowth, as reported previously24. The flexible nature of the neural interface (Supplementary Fig. 2) enables desirable mechanical properties when interfacing with soft peripheral nerves.

An additional biodegradable shape memory polymer (SMP) layer (~400 μm) consisting of PLLA-PTMC with a different copolymer ratio (70:30) is integrated with the device to facilitate implantation procedure by self-rolling around the peripheral nerve at body temperature (Supplementary Fig. 1). The copolymer ratio of 70:30 is adopted for PLLA-PTMC to ensure that the recovery temperature is adjusted to physiological temperature range25. The device in a rolled-up state fixed by the shape memory polymer appears in Fig. 2a, and the recovery from the temporarily flattened state (room temperature, in air) to the final self-curved state (37 oC, in phosphate buffered saline (PBS)) is given in Fig. 2b. The differential scanning calorimeter (DSC) measurement of the SMP indicates that the glass transition temperature (Ttrans) is approximately 36.14 °C, which confirms the shape memory effects at body temperature. Supplementary Mov. 1 illustrates the deployment of the device utilizing SMP on the intact nerve of Sprague-Dawley (SD) rats by applying PBS at 37 oC.

Fig. 2: Biodegradable, restorative and self-morphing neural interface.
figure 2

a Image of the biodegradable and restorative neural interface: front view (left) and side view (right). b Neural interface in the temporarily flattened state in air at room temperature (left) and in the permanent rolled-up state after immersion in the PBS (pH 7.4, 37 °C) solution (right). c, DSC analysis of the biodegradable shape memory polymer layer. d Representative electrochemical impedance spectra of the biodegradable Mo/Si recording electrode. e Electrochemical impedance at 1 kHz of the biodegradable Mo/Si recording electrode before and after immersion in PBS over 7 weeks, data are presented as mean ± s.d., n = 3 samples in each group. f Images collected at various stages of accelerated dissolution of Bio-Restor in PBS (pH 7.4, 60 °C). g Biodegradable Mo/Si electrode array implanted on the intact sciatic nerve. h Recorded signal traces using the Mo/Si electrode array on sciatic nerves during the flexion-extension movement of the hindlimb. The box plot presents the median (center line), lower quartile (lower border), upper quartile (upper border), maximum (upper whisker), and minimum (lower whisker), which are ≤ 1.5 times the interquartile range.

Figure 2d presents the representative electrochemical impedance spectra data for the Mo/Si electrodes in PBS at room temperature with an exposure area of ~1 mm2. The impedance at 1 kHz is 8.1 ± 0.1 kΩ, which is comparable to conventional gold (Au) electrodes with identical dimensions (Supplementary Fig. 3). Moreover, the impedance of Si/Mo electrodes demonstrates reasonable stability after soaking in PBS at 37 °C up to week 5, with an increase in impedance from 35.6 ± 9.8 kΩ (week 1) to 61.1 ± 3.9 kΩ (week 3), and 89.6 ± 7.7 kΩ (week 5) (Fig. 2e). Similar rise in impedance has been reported for non-degradable electrodes26,27. At week 7, a significant increase in impedance (451.9 ± 131.5 kΩ) is observed, likely resulting from the extensive dissolution of electrode materials. To evaluate the degradation characteristics of Bio-Restor, dissolution is performed in PBS solutions under accelerated conditions (pH 7.4, 60 °C), and the results are given in Fig. 2f. The Mg electrodes of the galvanic cell and Mo contacts without Si encapsulation dissolve within 2 days due to relatively rapid degradation. Mo contacts are therefore always coated with biodegradable polymer before implantation to extend the operational time frame. The subsequent disappearance of the FeMn galvanic electrodes and bilayer Mo/Si recording electrodes occurs after approximately 9 days. The degradation rate of Si is estimated to be approximately 4–7 times slower at 37 °C than at 60 °C28. Therefore, the lifespan of the recording electrodes at 37 °C is estimated to be around 5.5–8.5 weeks, aligning well with the results demonstrated in Fig. 2e. The polymeric substrate materials degrade through hydrolysis almost entirely after approximately 20 days.

To evaluate the recording performance of the Mo/Si electrodes, we implant the electrode array on the intact sciatic nerves of SD rats (Fig. 2g). A commercial neural cuff based on platinum (Pt) electrodes is also investigated as a reference (Supplementary Fig. 4). It is noted that the Mo/Si electrode array can easily wrap around the sciatic nerve due to the incorporation of SMP, enabling a seamless device/tissue interface and minimizing potential tissue damage. Evoked neural signals (compound action potential) is first investigated by applying electrical stimulation at the proximal side (Supplementary Fig. 5). Corresponding results are given in Supplementary Fig. 6 and 7, and the recorded signals are comparable between the cuff electrodes and the biodegradable electrodes (Supplementary Fig. 6). Additionally, reduced stimulation voltage results in reduced signal amplitude, and the evoked neural signals remain similar with inverted stimulation phase (Supplementary Fig. 7). Furthermore, electrical signals recorded on sciatic nerves of SD rats during periodic movement of the hindlimb exhibit a strong correlation with the flexion-extension movement, and representative results are shown in Fig. 2h. To eliminate the electromyography (EMG) artifact, we processed the data by referencing the recordings to the mean signal of all electrodes. The signals are comparable to the those obtained by the cuff electrodes (Supplementary Fig. 8). Collectively, these results demonstrate that the bioresorbable Mo/Si electrodes are reliable in recording neural signals on sciatic nerves. We then assess the feasibility of utilizing the biodegradable interface for monitoring nerve recovery progress and detection of neuroma to enable timely intervention.

Biodegradable neural interface for tracking nerve recovery

We integrate a wireless bio-potential recorder with the biodegradable neural interface to enable signal amplification, digitization, and wireless data transmission during the free movement of SD rats, as depicted in Fig. 3a. Figure 3b presents the specific circuit structure of the wireless circuit, data recording, and video capturing system. An analog front end (AFE) with chopper amplification technology is designed to acquire and convert the neural signal to digital data while maintaining low noise and low power consumption. The design was fabricated with 180 nm technology featuring a die size of 5 mm by 2.95 mm. The measured results demonstrate an integrated input-referred-noise of the low noise amplifier (LNA) of 0.79 μV from 1 Hz to 1 kHz, and a spurious-free dynamic range (SFDR) of 76 dB with a sinusoid input (93 Hz, 14 mVpp), as shown in Supplementary Fig. 9.

Fig. 3: Evaluation of nerve recovery status using Bio-Restor.
figure 3

a Image of a freely moving rat with an implanted Bio-Restor and a wireless module. The enlarged view shows the image of the circuit chip. b Schematic illustration of the wireless circuit, data recording and video capturing system. c Transected sciatic nerve with a 10-mm gap in SD rats. d Bio-Restor implanted at the injured site of nerve gap. e 3D and f cross-sectional micro-computer tomography images of the implanted Bio-Restor. Dotted orange lines indicate the position of connected wires to the multi-channel electrodes. g, Immunofluorescent images of the longitudinal section of the nerve segment with implanted Bio-Neuro at 3 weeks postimplantation. Top: immunohistochemical analysis demonstrating the leading edge (white dashed lines) of axonal growth. Yellow dashed lines indicate the position of neural electrodes. Bottom: enlarged view of the immunofluorescent images at 3 different locations. Immunohistochemical staining: axons (NF200, green), Schwann cells (S100, red), and nuclei (DAPI, blue). h Recorded signal traces using Bio-Neuro with rats walking on a treadmill at 3 weeks postimplantation. i Immunofluorescent images of the longitudinal section of the nerve segment with implanted Bio-Restor at 3 weeks postimplantation. Top: Immunohistochemical analysis demonstrating the leading edge (white dashed lines) of axonal outgrowth. Yellow dashed lines indicate the position of neural electrodes. Bottom: enlarged view of the immunofluorescent images at 3 different locations. j Recorded signal traces using Bio-Restor with rats walking on a treadmill at 3 weeks postimplantation. In g and i, experiments were repeated with n = 3 rats per condition.

To evaluate the capability of tracking nerve recovery in long-gap nerve injuries, we implant Bio-Restor to bridge 10-mm nerve defects in the sciatic nerves of SD rats (Supplementary Fig. 10), with a bipolar configuration to reduce electromyographic noise (Supplementary Fig. 11). The biodegradable neural interface without the galvanic cell is also investigated for comparison, referred to as the Bio-Neuro group (Supplementary Fig. 10). The images of implanted devices and implantation procedure are illustrated in Figs. 3c, d and Supplementary Fig. 12. The biodegradable neural interface resides at the injured site, routes subcutaneously up to the neck, and connects percutaneously to the wireless circuit for remote signal recording. Rats are able to move freely after device implantation (Supplementary Mov. 2). Figure 3e, f displays a micro-computer tomography (micro-CT) constructed 3D image and the cross-sectional view of the implanted device respectively. Wireless signal recording of free-moving rats on a treadmill is conducted at week 2 and week 3 (Supplementary Fig. 13), which represents the critical initial stage of nerve recovery. At week 3, the skin wound of SD rats at the implantation site has fully healed (Supplementary Fig. 14) and the retrieved biodegradable electrodes in the Bio-Restor remain intact (Supplementary Fig. 15). Supplementary Fig. 16 display the images of Hematoxylin and Eosin (H&E) staining of the regenerated axons in the Bio-Restor and Bio-Neuro groups at week 3, indicating no significant inflammation and thus desirable biocompatibility of the device.

Representative recorded signal traces of the Bio-Restor and Bio-Neuro groups and corresponding immunofluorescent images of longitudinal sections of regenerated nerve segments at 3 weeks postimplantation is shown in Fig. 3g–j. For signal processing, we utilize low-frequency band filters, as the measured electrical signals were primarily distributed in the range of 1 to 90 Hz, as illustrated in the frequency-time diagram in Supplementary Fig. 17. The electrical signals recorded using both low-frequency and high-frequency band filters are presented in Supplementary Fig. 18. During normal movement, we do not observe significant spikes or baseline modulation in the high-frequency range. The demonstration of distinct signals recorded in different electrode channels across the nerve defects (Fig. 3h) may indicate not significant EMG interference, as significant interference would typically lead to similar signals across all channels. Interestingly, we can observe a correlation of the leading edge of regenerated axons and the position of recording electrodes that are able to pick up signals. This is because electrodes located distally from regenerated axons are less likely to sense compound neural signals. Specifically, the electrode position is determined by the relative location of the recording electrodes and the footprints of electrodes left on the histological images (e.g., the creation of holes after tissue sectioning in Fig. 3g, i and Supplementary Fig. 16). The enlarged images in Fig. 3g, i demonstrate the representative morphology of regenerated axons (green fluorescent area), with the last one depicting the leading edge of axonal growth. In the Bio-Restor group, the front end of axons has reached around the position of electrode 7 (Fig. 3i), and all the electrode channels exhibit recorded signals (Fig. 3j). By contrast, the control group (Bio-Neuro group) demonstrates the leading edge of regenerated axon has only reached the position around electrode 5 (Fig. 3g), and only 4 channels recorded signals with significant amplitude (Fig. 3h). Similar correlation between histological and signal results are also observed at week 2 in the Bio-Neuro group (Supplementary Fig. 19). Additionally, we perform statistical analysis on axon density at week 3 following the method in the previous work24, and the results indicate a significant increase in axon density (p < 0.001)  in the Bio-Restor group than that of the Bio-Neuro group (Supplementary Fig. 20). Furthermore, we observe hindlimb movement in the Bio-Restor group at week 3 upon evoked stimulation on the proximal nerve segment, indicating muscle reinnervation has been achieved (Supplementary Mov. 4). By contrast, such hindlimb movement was not observed upon evoked stimulation in the Bio-Neuro group at week 3. The faster restoration of axonal growth in the Bio-Restor group is attributed to the electrical cues offered by the integrated galvanic cell, consistent with previous report24. These results suggest that the position of the leading edge of regenerated axons could be qualitatively tracked based on the electrode channels that are able to record significant signals, allowing real-time monitoring of regrowth rates and therefore potentially enabling timely intervention to ensure optimal recovery outcomes.

It is noted that for the signals recorded during treadmill walking, we did not reference recordings to the mean signal of all electrodes. Our intention is to investigate electrical signals on the injured nerve, where the signals from different channels may vary significantly, which can be utilized to qualitatively identify the leading edge of axonal growth. Referencing signals to the mean signals of all electrodes would not enable the capture of variations among different electrode channels. There is an inherent challenge in recording neural signals from freely-moving animals due to persistent EMG interference in the low-frequency range, even with reference electrodes and data filtering. The currently measured electrical data exhibits relatively large amplitude and low frequency, which may be attributed to the characteristics of the regenerated nerve segment. Compared to the non-regenerated segment, the regenerated segment may cause the electrodes in contact with it to detect increased activity, whether originating from the nerve, EMG signals, or movement. Future efforts to optimize electrode configurations and filtering algorithm may assist further reduction of EMG and movement interferences. Nevertheless, for the present study, our focus remains on detecting signal variations during nerve regrowth processes and neuroma presence (as will be shown later), with the expectation that moderate EMG interference will not significantly impact our analysis.

With the assistance of machine learning methods, we can further analyze the recorded electrical signals during treadmill walking to establish a correlation with gait parameters. These results can provide an important indicator of the recovery status, as nerve injury and regeneration often lead to changes in kinematics and spatiotemporal parameters of gait. Treadmill and Catwalk gait analysis is first conducted using traditional methods, which involve high-speed camera recordings from the lateral and bottom views respectively (Fig. 4a). The treadmill gait analysis includes the placement of markers on the hip, knee, ankle, and toe for subsequent DeepLabcut analysis (Supplementary Fig. 21), to obtain essential gait parameters such as joint mechanics, swing phase and stance phase. Specifically, we focus on ankle angle and the duration of the swing phase and stance phase. The stance phase is the period during which the foot is in contact with the ground (Supplementary Fig. 22a), while the swing phase is when the foot raises off the ground (Supplementary Fig. 22b) and swings forward in preparation for the next step. Both the Bio-Restor and Bio-Neuro groups at week 2 and week 3 are investigated, with the gait of normal rats served as the reference. The Bio-Restor group at week 3 exhibits significantly greater ankle angle (p < 0.05) and shorter swing time (p < 0.05) compared to the Bio-Neuro group at week 2. This is evident from the stick diagram in Fig. 4b and S23, ankle angle in terminal stance in Supplementary Fig. 24, and actual swing time in Fig. 4c. These findings indicate a better recovery in the Bio-Restor group at week 3. Moreover, a comparable pattern is noted in the Catwalk analysis, where the representative footprints of the injured hindlimb in the Bio-Restor group at week 3 appear more intact than those in the Bio-Neuro group at week 2 (Fig. 4d and S25). The associated sciatic functional index (SFI)24, calculated from these footprints, shows a significantly higher value (p < 0.01) in the Bio-Restor group at week 3 compared to the Bio-Neuro group at week 2, signifying improved motor functional recovery.

Fig. 4: Machine learning assisted evaluation of nerve recovery status through the analysis of swing phase based on recorded electrical signals.
figure 4

a Schematic illustration of conventional gait analysis methods including recording gait from the lateral sides on the treadmill (treadmill gait analysis) and foot prints underneath the runway (Catwalk gait analysis) by high-speed camera. b Kinematic stick diagram decompositions of hindlimb movement on a treadmill. c Statistical analysis of actual swing time of the Bio-Neuro and Bio-Restor groups at 2 and 3 weeks postimplantation. n = 4 rats in each group. d Representative hindlimb footprints for both injured and contralateral (healthy) sides of the Bio-Neuro group at 2 weeks postimplantation and the Bio-Restor group at 3 weeks postimplantation. e, Statistical analysis of sciatic functional index. n = 6 rats in each group. f Electrical signals recorded by the Bio-Neuro at week 2 during walking and resting states. g Convolutional neural networks (CNN) architecture used to evaluate the swing phase and stance phase based on recorded electrical signals and corresponding gait videos. h Representative results of raster plots of electrical signals, corresponding actual swing phase based on gait videos, and predicted swing phase by CNN architecture of the Bio-Restor group at week 3 postimplantation. i Confusion matrix of the prediction and ground truth of swing and stance phase of the Bio-Restor group at week 3. j Statistical analysis of the predicted swing time of the Bio-Neuro group at week 2 and the Bio-Restor group at week 3. n = 5 samples in each group. In c and e, One-way ANOVA followed by Turkey test was used for the statistical analysis (*P < 0.05, **P < 0.01, ***P < 0.001: P = 3.8 × 10-5, pairwise comparisons not shown on graphs are p > 0.05 (not significant)). In j, unpaired two-sided t-test was used for the statistical analysis. The box plot presents the median (center line), lower quartile (lower border), upper quartile (upper border), maximum (upper whisker) and minimum (lower whisker), which are ≤ 1.5 times the interquartile range.

Given the notable distinctions in gait results obtained through traditional methods between the Bio-Restor group at week 3 and the Bio-Neuro group at week 2, we proceed to analyze the recorded electrical signal traces of these two groups to establish correlations with gait parameters. This analysis aims to assess the recovery status based on electrical recording. Representative recorded signal traces of the Bio-Neuro group at week 2 are given in Fig. 4f and Supplementary Mov. 3, and a correlation is observed between the patterns of electrical signals and the phases of walking or resting. The recorded signals are fed into a Convolutional Neural Network (CNN) to perform swing and stance phase recognition (Fig. 4g). Electrical signal dataset was preprocessed before training the network to remove obvious artifacts resulting from movements to improve accuracy. Overlapping windows and oversampling technique are employed for data augmentation and sample balancing. Convolutional layers are applied for feature extraction from the raw acquired neural samples. The pooling and dropout layers are applied to avoid overfitting20. The representative result of the Bio-Restor group at week 3, including predicted swing phase, raster plot based on electrical recording, and actual swing phase obtained by camera recording, is shown in Fig. 4h. A reasonable classification accuracy in terms of swing (77%) and stance phase (85%) are achieved (Fig. 4i and S26). The classification accuracy of the Bio-Neuro group at week 2 is 72% and 79% for swing and stance phase, respectively (Supplementary Fig. 26). Furthermore, the predicted swing time indicating the recovery progress can be obtained through CNN analysis, and the results of the Bio-Restor group at week 3 and the Bio-Neuro group at week 2 are summarized in Fig. 4j. The results indicate a significant reduction of swing time for the Bio-Restor group at week 3 compared to that of the Bio-Neuro group at week 2 (p < 0.01). This trend is consistent with the gait analysis obtained by conventional methods (Fig. 4b–e). These results altogether suggest that the real-time electrical signals recorded by the bioresorbable interface can be utilized to remotely interrogate the leading edge of axonal growth and gait parameters, therefore predicting the recovery status of injured nerves without the need for camera-based gait analysis.

Biodegradable neural interface for early detection of neuroma

Neuroma often occurs with severe nerve injuries and amputations, and manifests as intricate clusters of disorganized axons encapsulated within fibrous scar tissues at the injured site29. Clinical studies have indicated that neuroma prevalence rates range from 40% to 70% following nerve transection30. The presence of neuromas could lead to the subsequent persistence of neuropathic pain and pose formidable obstacles to the effective implementation of prosthetic devices in the cases of amputation. Conventional diagnostic methods for neuroma utilize physical assessments, imaging technologies, and intraoperative electrodiagnostic tests. However, the obstacle arises from the challenge of enabling real-time and remote monitoring, thereby impeding early diagnosis. The biodegradable neural interface offers a promising solution to address these issues, allowing early detection of neuroma and timely therapeutic intervention, which is vital for achieving successful reinnervation and optimizing clinical outcomes. To evaluate the feasibility of neuroma detection, we create 10-mm nerve defects on the sciatic nerves. The Bio-Restor is placed at the proximal side of the sciatic nerve for direct signal recording while liberating the distal nerve stump to allow neuroma formation (Fig. 5a). Once neuroma is detected, timely intervention (e.g., surgical excision) is possible, enabling a better recovery (Fig. 5a).

Fig. 5: Early detection of neuroma using Bio-Restor.
figure 5

a Schematic illustration of the formation of neuroma, detection by Bio-Restor, and early intervention to achieve improved therapeutic outcomes. b Photographs, ultrasound and immunofluorescent images of neuroma at 2 weeks post injuries. c Photographs, ultrasound, and immunofluorescent images of neuroma at 3 weeks post injuries. d–f Representative recorded evoked signals by Bio-Restor of the control (freshly transected nerve, measured at the proximal end close to the transected site), neuroma at 2 weeks, and neuroma at 3 weeks respectively, under the electrical stimulation at the proximal side with a voltage amplitude of 1 V, 5 V and 5 V respectively. Stimulation frequency: 1 Hz, pulse duration: 10 ms. Purple square represents stimulation artifacts. g Statistical analysis of the amplitude of evoked neural signals. h–j Representative recorded electrical signals and associated time-frequency spectrograms of rats walking on a treadmill of the control (regenerated nerves at 3 weeks), neuroma at 2 weeks, and 3 weeks respectively. k Confusion matrix of the prediction and ground truth of the control, neuroma 2 W, and neuroma 3 W groups based on evoked neural signals. l Confusion matrix of the prediction and ground truth of the control, neuroma 2 W and neuroma 3 W groups based on electrical signals recorded during walking on a treadmill. In b, c, experiments were repeated with n = 3 rats per condition. In df, data are presented as mean ± s.d. In g, n = 10 measurement of 3 rats in each group, and one-way ANOVA followed by Dunnett’s T3 test was used for the statistical analysis (***P < 0.001: Control-2W: P = 5.8 × 10-5, Control_3W: P = 4.7 × 10-7, 2W-3W: P = 6.8 × 10-4). The box plot presents the median (center line), lower quartile (lower border), upper quartile (upper border), maximum (upper whisker), and minimum (lower whisker), which are ≤ 1.5 times the interquartile range.

At 2 weeks post nerve injuries, the formation of enlarged and bulbous nerve tissues are observed (optical images in Fig. 5b), indicating the occurrence of neuroma. Immunofluorescent staining of the nerve tissues reveals chaotic and disorganized axonal growth (displayed as the green fluorescent area in Fig. 5b) instead of the typical directional neurites seen in healthy nerves, further confirming the formation of neuroma. This group is referred as the Neuroma 2 W group. An increase in the size of neuromas is observed at 3 weeks post nerve injuries (Fig. 5c), referred to as the Neuroma 3 W group. Notably, identification of the occurrence of neuroma at such an early stage cannot be unambiguously determined with traditional ultrasound techniques (ultrasound images in Fig. 5b, c), and tissue harvesting is needed for morphological and immunohistochemical analysis to confirm its presence. In contrast, the biodegradable nerve interface is able to identify significantly different electrical signal patterns at week 2 and week 3, achieving early neuroma detection. Firstly, we investigate evoked neural signals in the presence of neuroma, and representative results are given in Fig. 5d–f. We study signal traces recorded by the Bio-Restor device in the control (freshly transected nerve, measured at the proximal end close to the transected site), neuroma 2 W and neuroma 3 W groups, upon electrical stimulation on the proximal site. The results show a significant reduction in amplitude with the presence of a neuroma, and an increase in neuroma size leads to a further decrease in amplitude (Fig. 5g). This is likely due to the fibrotic encapsulation of disorganized neural tissues, which greatly blocks electrical signals. Clinical studies have reported that poor recovery of neuroma-in-continuity is associated with significant decrease in the amplitude of nerve action potential during intraoperative measurement31, which further validates the evoked neural recording with neuroma formation in our work.

Furthermore, electrical signal traces recorded during treadmill walking are investigated. With the biodegradable neural interface, real-time monitoring can be accomplished remotely without the need of additional devices (e.g., stimulation electrodes). Representative results and associated time-frequency spectrograms are given in 5h–j, for the control, neuroma 2 W, and neuroma 3 W groups. Here, the regenerated nerves without neuroma formation at week 3 (Bio-Restor group at week 3) is adopted as the control group. The results indicate a decrease in the relative energy within the high-frequency band (approximately 15–50 Hz) is associated with the formation of the neuroma (neuroma 2 W and neuroma 3 W groups). This is likely caused by the presence of thick fibrotic tissues due to neuroma formation, similar to the mechanism that causes the decrease in the amplitude of evoked neural signals. The distinct patterns of electrical signals can be analyzed to enable early diagnosis of neuroma. With the assistance of machine learning analysis (CNN), a high-accuracy of classification can be accomplished and the results are given in Fig. 5k, l. The identification accuracy of neuroma formation at week 2 is 94% and 93%, based on signals recorded during evoked stimulation and treadmill walking respectively. In comparison, traditional ultrasound technique is unable to detect neuroma formation at such an early stage (Fig. 5b). Moreover, the identification accuracy for Neuroma 3 W is 92% and 96%, respectively, based on signals recorded during evoked stimulation and treadmill walking.

Due to the sensitivity of electrical signals to passivation, these findings indicate that the neural interface implanted around the injured nerve could provide a robust tool for identifying neuroma formation. In comparison to the traditional electrophysiological assessment performed on target muscles, which necessitates extensive time for reinnervation to elicit a visible response, directly examining electrical activity at the injury site allows for much earlier detection31. Overall, these results suggest that the biodegradable interface can provide essential information for early diagnosis of neuroma formation. Nevertheless, to address the concern of false positives or negatives in detection, rigorous testing and validation processes are needed in the future. Strategies include optimizing algorithms through the collection of a large and diverse datasets, and employing cross-validation techniques such as splitting the dataset into training and testing subsets to assess the robustness of the algorithm. Additionally, we can also strategically adjust parameter weights to minimize the missed detection rate.

Early detection of neuroma is crucial to enable timely intervention and enhance therapeutic efficacy. To demonstrate the importance of early diagnosis, we investigate the outcomes of intervention at different time points. Specifically, upon the detection of neuroma at 2 weeks post nerve injuries, we follow a standard clinical treatment, by performing excision of the neuroma and employing a nerve conduit to guide nerve regrowth (intervention at 2 W) (Fig. 6a). For comparison, the same intervention at a later stage, 5 weeks post nerve injuries (intervention at 5 W), is performed. The outcomes of gastrocnemius muscle atrophy and nerve conduction recovery at 12 weeks post-intervention are evaluated. Figure 6b presents the representative image of gastrocnemius muscle of both the injured and contralateral (healthy) sides, suggesting significant muscle atrophy with a later intervention (5 W). To quantitatively assess muscle atrophy, the muscle wet weight ratio is calculated by comparing the gastrocnemius muscles on the injured side with those on the contralateral healthy side, as shown in Fig. 6c. The results indicate that the muscle wet weight ratio of the intervention at week 2 is significantly greater than that of the intervention at week 5 (p < 0.05). Figure 6d displays the Masson trichrome staining of gastrocnemius muscles in the injured limbs, revealing notable reduction in collagenous deposits and larger average cross-sectional muscle fiber area in the early intervention (2 W) group. Statistical analysis of the cross-sectional area of muscle fibers demonstrates that the muscle fiber area in the early intervention group (2 W) are significantly greater than that of the later intervention group (5 W) (p < 0.001) (Fig. 6e). Moreover, nerve conduction recovery is examined by assessing compound muscle action potential (CMAP), which is recorded on the targeted gastrocnemius muscles upon the electrical stimulation of the proximal sciatic nerve stumps on the injured side. The representative results are illustrated in Fig. 6f. Statistical analysis reveals a significant improvement in CMAP amplitude in the early intervention group (2 W) compared to the later intervention group (5 W) (p < 0.05). In all, these results suggest the essential benefits of early neuroma detection and intervention, which can greatly facilitate nerve reinnervation and reduce muscle atrophy.

Fig. 6: Therapeutic outcomes of early intervention of neuroma 12 weeks postoperatively.
figure 6

a Photograph of the treatment of neuroma by excision and nerve guidance conduits. b Photographs of the gastrocnemius muscles of the contralateral (healthy, left) and injured side (right) with intervention at week 2 and week 5. c, Statistical analysis of the wet weight ratio of the gastrocnemius muscles of the injured hindlimb. d Masson’s trichrome staining images of the transverse sections of muscles of the injured hindlimb. e, Statistical analysis of the cross-sectional area of muscle fibers of the injured hindlimb. f Representative CMAP at the injured side. Electrical stimulation (1.0 mA, 1 Hz, 0.1 ms) is applied at the proximal nerve stumps. g Statistical analysis of CMAP amplitude at the injured side. In c, e and g, n = 3 rats per condition, and unpaired two-sided t-test was used for the statistical analysis (*P < 0.05, ***P < 0.001: P = 1.3 × 10-59). The box plot presents the median (center line), lower quartile (lower border), upper quartile (upper border), maximum (upper whisker), and minimum (lower whisker), which are ≤ 1.5 times the interquartile range.

Biodegradation of the neural interface

The essential feature of the proposed neural interface is that it is constructed with all biodegradable materials and can therefore be resorbed by the body over time after tracking nerve recovery, eliminating surgical complications for device retrieval and minimizing secondary damages to delicate nerve tissues. The active components including Mo/Si electrodes and the galvanic cell largely degrade after 7 weeks of implantation, and the remaining polymeric substrates degrades over a much longer time frame through hydrolysis (Fig. 7a). The morphology of the retrieved device, isolated from regenerated nerve tissues, is displayed in the inset of Fig. 7a, illustrating a gradual degradation process. The recorded electrical signals remain stable at week 2 and week 3, which provides a sufficient window for monitoring the critical stage of long-gap nerve injuries. The signal starts to decay at week 5 and completely lost the recording capability at week 7 (Fig. 7b). Moreover, biochemical analyses of the blood samples of SD rats with device implantation indicate the absence of physiological abnormalities 3 weeks postimplantation with a reference to the control group (Fig. 7c). Histological examinations of vital organs, including the heart, liver, spleen, lung, and kidney tissues, reveal no significant adverse effects compared to the control group, suggesting desirable biocompatibility of the biodegradable neural interface (Fig. 7d).

Fig. 7: Biodegradable and biocompatible characteristics of Bio-Restor.
figure 7

a Images of the Bio-Restor implanted at the injured sites of sciatic nerves in rodents at different stages of biodegradation over a time frame of 17 weeks. Inset: retrieved device isolated from regenerated nerve tissues. b Representative recorded signal traces and corresponding time-frequency spectrograms at different stages during the course of biodegradation. c Analysis of blood chemistry and blood counts of rats of the control (no device implantation) and Bio-Restor groups at 3 weeks postimplantation. n = 3 rats in each group, data are presented as mean ± s.d. ALT (Aminotransferase); TBIL (Total Bilirubin); BUN (Blood Urea Nitrogen); CR (Creatinine). Dash lines represent the normal range. d H&E-stained histological sections of heart, liver, spleen, lung, and kidney of the control and Bio-Restor groups at 3 weeks postimplantation. Experiments were repeated with n = 3 rats per condition.

Discussions

We introduce a biodegradable and restorative peripheral neural interface that not only supports nerve regrowth but also enables real-time electrical recording at the critical early stage following neuropathic injuries. The enhanced regeneration effect is attributed to the electrical cues generated by the galvanic cell. Electric field has been proved to promote nerve repair. The beneficial effects could results from accelerated Wallerian degeneration32, and prompt and continuous upregulation of regeneration-associated genes33,34, therefore facilitating more rapid axonal regrowth. The electric field strength along the conduit is expected to range from 20 to 250 mV/mm24, consistent with the electric field needed to promote nerve regeneration reported in previous research35,36. While the therapeutic effects for tissue regeneration and pain management based on optoelectronic or electrical stimulation have been demonstrated previously24,37,38, there is limited prior work on biodegradable nerve interfaces that are capable of signal recording. We now showcase the capability of remote monitoring nerve recovery status by tracking the leading edge of axonal growth and the prediction of gait parameters. Furthermore, early detection of the formation of traumatic neuroma is accomplished, which allows timely intervention to significantly improve therapeutic outcomes. The device platform also has potential applications for amputees to facilitate TMR procedures, which can assist reinnervation, identify potential neuroma formation, and therefore facilitate earlier and more effective prosthetic integration. The neural interface is constructed with bioresorbable materials that can degrade into harmless byproducts after the monitoring period. The proposed materials strategies and device schemes provide new avenues to interrogate previously unexplored electrical neural signals for real-time monitoring and identification of abnormities following nerve injuries, which could greatly improve the efficacy of treating long-segment nerve injuries. Further optimization of electrode and encapsulation materials can customize degradation rates and extend the working timeframe. For example, substituting the bioactive components of the galvanic cells with Mg alloys or Zn, and utilizing electrodes with greater thickness and encapsulation materials with slower dissolution rates will extend the life time of the neural interface. Moreover, employing triggered degradable materials presents an alternative method for precise control of the degradation period. These material strategies have the potential to prolong the operational lifespan into the later stages of nerve recovery, allowing devices to align with an individual’s unique healing trajectory. Additionally, the implantable wires can also be designed to be biodegradable in the future39, negating the need for surgical removal and reducing the risk of complications. Additionally, optimized machine learning algorithms can be designed to analyze the characteristics of recorded signals at various stages and evaluate the device’s stability. By leveraging this comprehensive data, the algorithm can be enhanced to predict outcomes effectively throughout different phases of nerve recovery. The increase of the number of channels can enhance the spatial resolution of input signals, allowing for the capture of more detailed and precise neural activity. This, in turn, offers deeper insights into the underlying neural processes at play. The incorporation of neural modulation components can facilitate achieving closed-loop system for enhanced recovery and pain management. Collectively, the research offers essential insights into examining electrical activity in the context of post-traumatic nerve injuries or neural disorders, which can uncover crucial information that can optimize treatment protocols, ultimately leading to maximized clinical outcomes and enhanced patient care.

Methods

Fabrication of biodegradable and restorative neural interface

Polymer films (70:30 PLLA-PTMC of SMP: 440 μm, 10 × 5 mm and 60:40 PLLA-PTMC: 30 μm, 10 × 10 mm) were fabricated by dissolving PLLA-PTMC (Jinan Daigang Biomaterial Co. Ltd., China) into trichloromethane (CHCl3) (Beijing Tongguang Chemical Co. Ltd., China) with a weight-to-volume (w/v) ratio of 1:10. The solutions were then drop-casted and cured for 12 hours at 4 °C to prevent the formation of bubbles. The SMP was then applied onto a needle (diameter, 1 mm) and subjected to heat treatment in an oven at 80 °C for a duration of 30 minutes. This process allowed the polymer to be shaped into a tubular structure. Subsequently, we reopened the tubular structure to a planar form at room temperature, and laminates it with the PLLA-PTMC (60:40) substrate using a minimal amount of CHCl3 solution.

Mo recording electrode array (300 nm, 0.3 × 5 mm) was deposited on the PLLA-PTMC layer material through a patterned shadow mask in a magnetron sputter (Beijing Zhongjingkeyi Co. Ltd., China) with a deposition speed of 1.5 Å s-1 (370 V, 0.3 A). Mg thin-film anodes (3.5 μm, 1.5 × 5 mm) and FeMn alloy cathodes (1.5 μm, 1.5 × 5 mm) were deposited on two sides of the bilayer material using magnetron sputter system through patterned shadow masks at a deposition speed of 1.8 Å s-1 (280 V, 0.23 A) and 1.1 Å s-1 (340 V, 0.25 A), respectively. The FeMn alloy is deposited by a customized target with 30 wt % Mn. Highly doped monocrystalline Si membranes (2 μm) were obtained using a silicon on insulator (SOI) wafer (top Si layer: conductivity 0.1–2 × 104 S·m-1, 2 μm). The top Si layer was patterned using a lithographic process followed by reactive ion etching with sulfur hexafluoride gas (SF6). Subsequently, Si membranes on top of the SOI wafer were released by hydrofluoric acid (49 % HF, ACS grade, Aladdin). Polydimethylsiloxane (Dow Corning Sylgard 184 kit, 1:10 weight ratio) stamps were applied to transfer the freestanding Si membranes onto the PLLA-PTMC substrate with the Mo electrode array to achieve Mo/Si bilayer recording electrodes.

Characterization of shape memory performance, electrochemical properties, and biodegradability

Differential scanning calorimetry (DSC) (Naichi Co. Ltd., Germany) was employed to determine the thermal properties of SMP. Nitrogen gas was used as a purging gas at a flow rate of 20 mL/min. The sample underwent an initial heating process, where it was heated from 0 to 100 °C at a rate of 10 °C/min and held at 100 °C for 2 minutes. Subsequently, it was cooled down to 0 °C at a cooling rate of 10 °C/min. This first heating process aimed to eliminate any thermal history effects arising from different materials. Following this, a second heating process was conducted to generate the DSC curve. The glass transition temperature (Tg) was determined from the second heating run.

The electrochemical impedance of the Mo/Si bilayer recording electrodes were investigated using a Gamry Interface 1000E Potentiostat. The measurements were conducted at room temperature in PBS. A three-electrode configuration was employed, with the Si/Mo recording electrodes serving as the working electrode, Ag/AgCl as the reference electrode, and Pt as the counter electrode. The electrochemical impedance spectroscopy (EIS) data were acquired by sweeping the frequency from 10 Hz to 0.1 MHz, which is relevant for neural sensing. A sinusoidal AC excitation signal with an amplitude of 5 mV and a bias of 0 V relative to the open-circuit voltage was applied for the measurement.

To evaluate degradation, the device was immersed in PBS (replaced every day) at 60 °C in a water bath. Time-series images were captured at various stages using an optical microscope.

Signal recording under evoked conditions

To evaluate the performance of neural recording, the biodegradable Mo/Si electrode array and commercial Pt cuff electrodes (Kedou (Suzhou) Brain-computer Technology Co. Ltd, China) were investigated on intact sciatic nerves of SD rats (n = 3 per condition). All trials were performed with adult female SD rats (220-250 g). Surgery was performed under 1.5% isoflurane (RWD Life Science Co., Shenzhen, China) anesthesia. Under sterile conditions, the right hindlimb was dissected to expose the sciatic nerve. To deploy the biodegradable Mo/Si electrode array to the sciatic nerve, a small amount of PBS at 37 °C was used to induce self-rolling of the SMP, ensuring conformal contact with the nerve tissues. Evoked neural signals were recorded upon the application of electrical stimulation (frequency:1 Hz, pulse duration: 10 ms, amplitude: ± 1 and 0.25 V) at the proximal side of the sciatic nerve. Signal traces were also recorded during the flexion-extension movement of the hindlimb.

Surgical procedure of device implantation

Long-gap sciatic nerve injuries: Twenty-four Female SD rats (220–250 g) were anesthetized using 3% pentobarbital sodium solutions (30 mg per kg body weight). The right sciatic nerve was exposed and the surrounding connective tissue was isolated. A 10-mm sciatic nerve defect was created proximal to the branch of the sciatic nerve using a microsurgery scissor. Bio-Restor or Bio-Neuro was connected to the nerve defect area and sutured using 8-0 polypropylene sutures (n = 6 per condition). The receiver routes subcutaneously along the spine to the neck. Finally, muscle layers and the skin were closed with 4-0 silk sutures.

Creation of traumatic neuroma: Eight female SD rats (220–250 g) were anesthetized using 3% pentobarbital sodium solution (30 mg per kg body weight). The right sciatic nerve was exposed, and the surrounding connective tissue was isolated. The nerve was cut 5 mm proximal to the sciatic nerve branch. The proximal nerve stump was enveloped with Bio-Restor and secured to the nearby muscle with 10-0 sutures. The distal nerve stump was folded back and sutured beneath the muscle layer with 10-0 sutures. The wound was then closed in successive layers. The formation of neuroma was examined at 2 weeks and 3 weeks postimplantation by optical observation, ultrasound, immunofluorescent staining (n = 3 per condition).

Intervention of traumatic neuroma: At 2 or 5 weeks following the primary surgery to create the neuroma, a secondary surgery was performed to repair the neuroma on the right sciatic nerve (n = 3 per condition). The base around the proximal nerve was meticulously dissected to examine the enlarged neuroma formation. The distal nerve was isolated from the muscle. The proximal neuroma and the scared area of the distal nerve were excised to reveal the fresh nerve stumps and create a 10 mm nerve gap. 8-0 sutures were utilized to connect the PLLA-PTMC conduit to the nerve defect region. Finally, the wound was closed layer by layer.

Gait analysis

Treadmill gait analysis: A customized treadmill setup and high-speed camera from the lateral sides was utilized to evaluate the walking patterns of rodents of the Bio-Restor and Bio-Neuro groups at week 2 and week 3 postimplantation (n = 6 per condition). Gait analysis was performed on normal rats for comparison. Markers were placed on the hip, knee, ankle, and toe of the rats as reference points for training of the algorithm. The gait parameters including joint angle, swing time, and stance time were extracted from the acquired video stream. A deep neural network (DeepLabcut), consisting of a pre-trained ResNet-50 backbone for feature extracting and deconvolutional layers for upsampling40, is employed to identify the locations of the reference points, with less than 5 pixels average error. Joint angles can be calculated from the coordinates of the reference points, and the swing and stance time can be measured by the toe height curve.

CatWalk gait analysis: CatWalk gait analysis was performed for the Bio-Restor and Bio-Neuro groups at week 2 and week 3 after surgery (n = 5 per condition). Briefly, walking tracks were recorded by a high-speed camera located beneath the runway using the CatWalk XT 10.6 gait analysis system (Noldus Information Technology, Wageningen, The Netherlands). Information about the foot contact area, walking posture, and 3D plantar pressure distribution was collected at the same time. At least three identifiable runs were performed for each rat in each measurement. Finally, SFI was calculated using Noldus Catwalk Analysis software. SFI was calculated using the formula outlined previously24:

$${{\rm{SFI}}}=\frac{109.5{{\rm{(ETS-NTS)}}}}{{{\rm{NTS}}}}-\frac{38.3\left({{\rm{EPL}}}-{{\rm{NPL}}}\right)}{{{\rm{NPL}}}}+\frac{13.3\left({{\rm{EIT}}}-{{\rm{NIT}}}\right)}{{{\rm{NIT}}}}-8.8$$
(1)

where ETS represents the experimental toe spread (the distance between the first and fifth toes), while NTS denotes the normal toe spread. EIT is the experimental intertoe spread (the distance between the second and fourth toes), and NIT is the normal intertoe spread. EPL denotes the experimental print length, and NPL denotes the normal print length.

In vivo data recording

In vivo signals were recorded for SD rats walking on a treadmill (n = 3 per condition). Evoked neural signals from the neuroma were recorded upon the application of electrical stimulation (duration: 50 μs, frequency: single pulse, amplitude: 5 V) to the proximal side of the sciatic nerve (n = 3 per condition).

Wireless circuit for signal recording and data transmission

The wireless circuit (20 × 20 × 4 mm) is based on a customized 16-channel analog-front-end (AFE) chip with flexible channel configurations. The nRF5340 Micro Control Unit (MCU) from Nordic Semiconductor is selected to drive the AFE and implement the BLE 5 protocol. The BLE 5 protocol with Data Length Extension (DLE) enables the real-time transmission of the collected electrical data. The entire wireless circuit consumes ~50 mW of power, which can be supplied by a 100 mAh battery (20 × 20 × 5 mm) with a lifespan of 9 hours. A 120-fps camera enables global shutter, preventing the potential image distortion when the experimental subjects are moving fast.

Data analysis of recorded signals

Signal processing: A Butterworth high-pass filter at 1 Hz is used to remove low-frequency motion artifact noise, while a low-pass filter with a 90 Hz cut-off frequency removes high-frequency environmental noise41. An IIR notch filter addresses industrial frequency noise and its harmonics at 50 Hz and its multiples (100 Hz, 150 Hz, 200 Hz). Wavelet denoising is also applied for signal processing and noise reduction.

Data decoding for gait prediction: In order to decode the neural signals of treadmill walking to establish a correlation with gait parameters, a 1D-CNN model is proposed to classify the acquired neural signals into two classes, including the swing phase and the stance phase. Based on gait patterns obtained by a high-speed camera, the electrical signal dataset was preprocessed before training the networks to remove the artifacts resulted from movements and muscle activities to improve accuracy. Overlapping windows and oversampling techniques were employed for data augmentation and sample balancing. We adjusted the length of the window and the overlapping ratio to test the performance of the classifier. The electrical signal dataset is divided into three subsets, 80% data for training, 10% data for validation, and 10% data for testing after preprocessing and normalization. The swing phase and stance phase can be identified from the electrical signal segment using 1D-CNN networks. Subsequently, the swing time and stance time can be calculated. The raster plot is achieved by drawing vertical lines at the corresponding time point to represent neural firing events.

Data decoding for neuroma identification: In order to decode the neural signals as well as to analyze neuroma growth, a 1D-CNN model was applied. The neural signal dataset of control, neuroma (2 W) and neuroma (3 W) was preprocessed before training the network to remove potential artifacts to improve the signal-to-noise ratio using the aforementioned methods. Neural signals collected in two different scenarios, including electrical stimulation and treadmill walking, were used for training and testing the network to comprehensively demonstrate the reliability of the results. After preprocessing and normalization, the electrical signals were segmented into non-overlapping 2-second segments, with 80% used as the training set, 10% as the validation set, and 10% as the test set. The 1D-CNN model can identify neural signal segments as either without neuroma (control), neuroma (2 W), or neuroma (3 W).

Histological and Immunohistochemical evaluation

At 2 and 3 weeks postimplantation, rats were euthanized with carbon dioxide gas (n = 3 per condition). Sciatic nerve tissue and devices were explanted and fixed in 4% paraformaldehyde for 48 hours. A frozen slicer (Leica, Wetzlar, Germany) was used to cut nerve tissues into 10-μm sections longitudinally. H&E and immunofluorescence staining were performed respectively. The following primary and second antibodies were used: Mouse anti-Neurofilament 200 antibody (N0142, Sigma, 1:300), Rabbit anti-S100 antibody (ab52642, Abcam, 1:200), Goat Anti-Mouse IgG H&L (Alexa Fluor® 488, ab150117, Abcam, 1:200) and Goat Anti-Rabbit IgG H&L (Alexa Fluor® 594, ab150084; Abcam, 1:200). Finally, the stained sections were scanned using a slide scanning system (3DHISTECH, Budapest, Hungary). All assessments were conducted in triplicate for each animal samples.

Evaluation of gastrocnemius muscles

Three SD rats from each group were chosen for assessment at 12 weeks post-surgery to analyze their electrophysiological response, wet weight ratio, and muscle fiber area in the gastrocnemius muscle. Motor nerve function was evaluated using a multi-channel physiological signal recording and processing system (RM6240EC, Chengyi). Under anesthesia (1% sodium phenobarbital), the sciatic nerves on the injured side were exposed. Electrical stimulation (3.0 mA, 1 Hz) was administered to the proximal nerve stumps, and CMAPs were recorded from the gastrocnemius muscle to assess motor nerve function. The peak amplitudes of CMAPs were measured and compared between the different groups. Following the recordings, the gastrocnemius muscles from both the injured and unharmed hindlimbs were excised and weighed to determine the wet weight ratio. The muscles were then fixed (4% paraformaldehyde, 12 hours), processed into thin paraffin sections (5 μm in thickness), and subjected to Masson staining for analysis. The stained sections were examined using a slide scanning system (3DHISTECH, Budapest, Hungary), and the cross-sectional area of the gastrocnemius fibers was quantified using Image-Pro Plus 6.0 software.

Micro-CT and ultrasonography

Micro CT of anesthetized rats was performed on a CT system (Lenovo Co. Ltd., China). Images were acquired with 70 kV, and the projection data were reconstructed with the Workstation. Ultrasound examination was performed at 2 and 3 weeks after the neuroma creation surgery. Rats were anesthetized with isoflurane, and the right hindlimb was shaved to fully expose the skin. High-frequency ultrasound equipment with a 21-to 44-MHz linear array probe (MX4000, Vevo 3100, Visualsonics, Canada) imaged the proximal stump of the sciatic nerve and measured the diameter of the enlarged nerve tissues.

In vivo biodegradation studies

For in vivo biodegradability evaluation, Bio-Restor was implanted in rats, which were subsequently exposed at 5, 7, and 17 weeks post-implantation. The corresponding Bio-Restor samples were then extracted, and photographs were taken to study the biodegradation behavior in vivo.

Assessment of hematology and blood chemistry

At 3 weeks postimplantation, rats were deeply anaesthetized and peripheral blood was collected from the orbital vein using vacuum blood collection tubes containing procoagulant and separator gel (QX0028, Servicebio). The blood samples were kept at room temperature for 2 hours, and then the serum was separated in a centrifuge speed at 1509.3 g for 15 minutes at 4 °C. Biochemical and electrolyte tests were performed by Servicebio Biotechnology Co. LTD.

Statistical analysis

Statistical analysis was performed using GraphPad Prism (version 6.0) for two-group comparisons with an unpaired t-test, and SPSS software (version 23.0) for multi-group comparisons using one-way ANOVA followed by Turkey test or Dunnett’s T3. Significance levels were set at *P < 0.05, **P < 0.01, ***P < 0.001, pairwise comparisons not shown on graphs were p > 0.05 (not significant). Data are presented as mean ± standard deviation.

Reporting summary

Further information on research design is available in the Nature Portfolio Reporting Summary linked to this article.